Literature DB >> 26520739

Luminescence imaging of water during proton-beam irradiation for range estimation.

Seiichi Yamamoto1, Toshiyuki Toshito2, Satoshi Okumura1, Masataka Komori1.   

Abstract

PURPOSE: Proton therapy has the ability to selectively deliver a dose to the target tumor, so the dose distribution should be accurately measured by a precise and efficient method. The authors found that luminescence was emitted from water during proton irradiation and conjectured that this phenomenon could be used for estimating the dose distribution.
METHODS: To achieve more accurate dose distribution, the authors set water phantoms on a table with a spot scanning proton therapy system and measured the luminescence images of these phantoms with a high-sensitivity, cooled charge coupled device camera during proton-beam irradiation. The authors imaged the phantoms of pure water, fluorescein solution, and an acrylic block.
RESULTS: The luminescence images of water phantoms taken during proton-beam irradiation showed clear Bragg peaks, and the measured proton ranges from the images were almost the same as those obtained with an ionization chamber. Furthermore, the image of the pure-water phantom showed almost the same distribution as the tap-water phantom, indicating that the luminescence image was not related to impurities in the water. The luminescence image of the fluorescein solution had ∼3 times higher intensity than water, with the same proton range as that of water. The luminescence image of the acrylic phantom had a 14.5% shorter proton range than that of water; the proton range in the acrylic phantom generally matched the calculated value. The luminescence images of the tap-water phantom during proton irradiation could be obtained in less than 2 s.
CONCLUSIONS: Luminescence imaging during proton-beam irradiation is promising as an effective method for range estimation in proton therapy.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26520739     DOI: 10.1118/1.4932630

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  7 in total

Review 1.  Innovations in Nuclear Imaging Instrumentation: Cerenkov Imaging.

Authors:  Ryo Tamura; Edwin C Pratt; Jan Grimm
Journal:  Semin Nucl Med       Date:  2018-03-16       Impact factor: 4.446

Review 2.  Photon counting detectors and their applications ranging from particle physics experiments to environmental radiation monitoring and medical imaging.

Authors:  Ryosuke Ota
Journal:  Radiol Phys Technol       Date:  2021-03-19

3.  Estimation of the fractions of luminescence of water at higher energy than Cerenkov-light threshold for various types of radiation.

Authors:  Yoshiyuki Hirano; Seiichi Yamamoto
Journal:  J Biomed Opt       Date:  2019-06       Impact factor: 3.170

4.  Optical imaging of muons.

Authors:  Seiichi Yamamoto; Kazuhiko Ninomiya; Naritoshi Kawamura; Yoshiyuki Hirano
Journal:  Sci Rep       Date:  2020-11-26       Impact factor: 4.379

5.  Enhancement of the ionoacoustic effect through ultrasound and photoacoustic contrast agents.

Authors:  Julie Lascaud; Pratik Dash; Matthias Würl; Hans-Peter Wieser; Benjamin Wollant; Ronaldo Kalunga; Walter Assmann; Dirk-André Clevert; Alfredo Ferrari; Paola Sala; Alessandro Stuart Savoia; Katia Parodi
Journal:  Sci Rep       Date:  2021-02-01       Impact factor: 4.379

6.  Increase in the intensity of an optical signal with fluorescein during proton and carbon-ion irradiation.

Authors:  Seiichi Yamamoto; Takuya Yabe; Takashi Akagi
Journal:  J Appl Clin Med Phys       Date:  2021-06-14       Impact factor: 2.102

7.  Detection of luminescence of radicals from TiO2 plate during alpha particle irradiation.

Authors:  Seiichi Yamamoto
Journal:  J Biomed Opt       Date:  2020-09       Impact factor: 3.170

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.