A candidate reference method for the determination of uric acid in serum based on high performance liquid chromatography, compared with an isotope dilution-gas chromatography-mass spectrometer method.

R Kock1, B Delvoux, U Tillmanns, H Greiling.   

Abstract

A method based on isocratic high performance liquid chromatography (HPLC) with UV detection at 292 nm is proposed as a candidate reference method for the determination of uric acid. Data obtained by this method are compared with those from an isotope dilution-gas chromatography-mass spectrometric method (ID-GC-MS), using [1,3-15N2]uric acid as internal standard and selected mass detection at m/z = 456 and m/z = 458. The inaccuracy of the ID-GC-MS method is maximally 0.4% for NBS-SRM-909 control sera with a concentration of 483 mumol/l. The coefficient of variation between days is 0.26%-0.80% and 0.37-0.90% for 14 control sera from other suppliers. The maximum bias of the HPLC method is 0.6%, and the coefficient of variation between days is 0.31%-0.65% for NBS-SRM-909 control sera. The coefficient of variation between days for the other 14 control sera tested is 0.35%-0.66%. Comparison of the HPLC method with the reference ID-GC-MS method resulted in a coefficient of correlation of r = 0.9998 (n = 14). The concentration of uric acid in the tested control sera ranged from 160 to 624 mumol/l.

Entities:  

Mesh:

Substances:

Year:  1989        PMID: 2651552     DOI: 10.1515/cclm.1989.27.3.157

Source DB:  PubMed          Journal:  J Clin Chem Clin Biochem        ISSN: 0340-076X


  1 in total

1.  Assessment of the trueness and inter-laboratory precision of routine uric acid assays using 4 frozen pooled serum samples measured by the Japan Society of Clinical Chemistry's HPLC method.

Authors:  Yaping Jiang; Ou Liu; Guobing Xu
Journal:  Ann Lab Med       Date:  2014-02-13       Impact factor: 3.464

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.