C K Egan1, S D M Jacques1,2, M D Wilson3, M C Veale3, P Seller3, A M Beale2,4, R A D Pattrick5, P J Withers1, R J Cernik1.
Abstract
We report the development of laboratory based hyperspectral X-ray computed tomography which allows the internal elemental chemistry of an object to be reconstructed and visualised in three dimensions. The method employs a spectroscopic X-ray imaging detector with sufficient energy resolution to distinguish individual elemental absorption edges. Elemental distributions can then be made by K-edge subtraction, or alternatively by voxel-wise spectral fitting to give relative atomic concentrations. We demonstrate its application to two material systems: studying the distribution of catalyst material on porous substrates for industrial scale chemical processing; and mapping of minerals and inclusion phases inside a mineralised ore sample. The method makes use of a standard laboratory X-ray source with measurement times similar to that required for conventional computed tomography.Entities:
Year: 2015 PMID: 26514938 PMCID: PMC4626840 DOI: 10.1038/srep15979
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Figure 1Hyperspectral XCT: A regular area detector in a commercially available microtomography scanner is replaced with a spectroscopic imaging detector with a high energy resolution.
The absorbed X-ray spectrum in each pixel can now be measured. A geometric magnification can be defined by moving the sample closer to the X-ray source enabling micrometer scale resolution. By recording projections at different angles of sample rotation, a 3D dataset of the sample object can be mathematically reconstructed. Each voxel now contains an absorption spectrum as opposed to a grayscale value. Step-changes in the absorption spectrum signify the position of absorption edges, which can be used to map the distribution of chemical elements inside the object with micron scale resolution.
Figure 2Pd metal distribution on a porous alumina catalyst body.
(a) Typical voxel spectrum from the sample showing a distinct step-change in attenuation at 24.3 keV, corresponding to the Pd K-edge. (b) Spectral fitting to the absorption spectrum. Two linear functions are least squares fitted above and below the edge. These are evaluated at the known edge position and subtracted to obtain Δμ0. By repeating this process for every voxel spectrum, the relative Pd metal concentration can be visualised in 3D. (c) and (d) Vertical and horizontal orthoslices of Δμ0. (e) 3D exterior visualisation of Δμ0.
Figure 33D distribution of mineral phases in a mineralised ore sample from a gold-rich hydrothermal vein.
(a) Grayscale tomographic slice through the sample created by integrating over the full spectral range. (b) Voxel spectra showing Au and Pb K-edges. (c) Voxel spectra from quartz, pyrite and chalcopyrite minerals. (d) Vertical slice through the sample with Au (blue) and Pb (red) containing voxels segmented and coloured. (e) 3D visualisations of the distribution of mineral phases in the sample.