Literature DB >> 26508129

Histological examination of vascular damage caused by stent retriever thrombectomy devices.

Daisuke Arai1, Akira Ishii1, Hideo Chihara1, Hiroyuki Ikeda1, Susumu Miyamoto1.   

Abstract

BACKGROUND AND OBJECTIVES: Although the recently marketed stent retriever thrombectomy devices have demonstrated a high recanalization rate and favorable clinical outcomes, there is a concern about the risks of intimal injuries when pulling out the stent in the unfolded position. In this study, the Solitaire Flow Restoration System and the Trevo retriever were used in a histopathological comparison of vascular injuries caused by stent retriever thrombectomy devices.
METHODS: Rabbit carotid arteries were used in the experiments with stent retriever thrombectomy devices. Carotid artery samples were harvested either 1 or 2 weeks postoperatively for histological examination.
RESULTS: Histological changes caused by the use of stent retriever thrombectomy devices were observed from the intimal to medial layers. With the Solitaire FR 4 mm, intimal and medial thickening was observed 1 week postoperatively, and progression of intimal thickening was observed 2 weeks postoperatively. The extent of intimal thickening tended to be greater with the Solitaire FR 6 mm than with the Solitaire FR 4 mm, but this difference was not significant. Compared with the Solitaire FR 4 mm, the Trevo had a significantly smaller area of intimal thickening.
CONCLUSIONS: Although there are some differences among devices, results from this study indicate that stent retriever thrombectomy devices induce vascular damage that extends to the medial layer. Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://www.bmj.com/company/products-services/rights-and-licensing/

Entities:  

Keywords:  Intervention; Stroke; Thrombectomy

Mesh:

Year:  2015        PMID: 26508129     DOI: 10.1136/neurintsurg-2015-011968

Source DB:  PubMed          Journal:  J Neurointerv Surg        ISSN: 1759-8478            Impact factor:   5.836


  20 in total

1.  Biomechanics and hemodynamics of stent-retrievers.

Authors:  Anna Luisa Kühn; Zeynep Vardar; Afif Kraitem; Robert M King; Vania Anagnostakou; Ajit S Puri; Matthew J Gounis
Journal:  J Cereb Blood Flow Metab       Date:  2020-05-19       Impact factor: 6.200

2.  Mechanical Thrombectomy with the Embolus Retriever with Interlinked Cages in Acute Ischemic Stroke: ERIC, the New Boy in the Class.

Authors:  H Steglich-Arnholm; D Kondziella; A Wagner; M E Cronqvist; K Hansen; T C Truelsen; L-H Krarup; J L S Højgaard; S Taudorf; H K Iversen; D W Krieger; M Holtmannspötter
Journal:  AJNR Am J Neuroradiol       Date:  2017-05-11       Impact factor: 3.825

3.  Increased Number of Passes and Double Stent Retriever Technique Induces Cumulative Injury on Arterial Wall After Mechanical Thrombectomy in a Swine Model.

Authors:  David Hernández; José Luis Cuevas; Laura Ludovica Gramegna; Manuel Requena; Carlos Piñana; Marta de Dios; Pilar Coscojuela; Marielle Esteves; Jiahui Li; Alberto Gil; Marc Ribó; Alejandro Tomasello
Journal:  Transl Stroke Res       Date:  2022-06-08       Impact factor: 6.829

4.  First pass effect vs multiple passes complete reperfusion: A retrospective study.

Authors:  Ameer E Hassan; Mahmoud Dibas; Amrou Sarraj; Sherief Ghozy; Amr Ehab El-Qushayri; Adam A Dmytriw; Wondwossen G Tekle
Journal:  Neuroradiol J       Date:  2021-08-31

5.  Relationship between the first pass effect and the platelet-lymphocyte ratio in acute ischemic stroke.

Authors:  Orkun Sarioglu; Ahmet Ergin Capar; Demet Funda Bas Sokmez; Pelin Topkaya; Umit Belet
Journal:  Interv Neuroradiol       Date:  2020-11-25       Impact factor: 1.764

6.  Comparison of Subacute Vascular Damage Caused by ADAPT versus Stent Retriever Devices after Thrombectomy in Acute Ischemic Stroke: Histological and Ultrastructural Study in an Animal Model.

Authors:  Simone Peschillo; Alejandro Tomasello; Francesco Diana; David Hernandez; Giada Toccaceli; Marta Rosal-Fontana; Marielle Esteves Coelho; Paolo Missori
Journal:  Interv Neurol       Date:  2018-08-31

7.  Favorable first-pass recanalization rates with NeVa™ thrombectomy device in acute stroke patients: Initial clinical experience.

Authors:  Cetin K Akpinar; Atilla O Ozdemir; Erdem Gurkas; Adnan B Bilgic; Ozlem Aykac; Yusuf Inanc; Semih Giray
Journal:  Interv Neuroradiol       Date:  2020-07-02       Impact factor: 1.610

8.  Understanding the Radial Force of Stroke Thrombectomy Devices to Minimize Vessel Wall Injury: Mechanical Bench Testing of the Radial Force Generated by a Novel Braided Thrombectomy Assist Device Compared to Laser-Cut Stent Retrievers in Simulated MCA Vessel Diameters.

Authors:  Jeffrey M Katz; Abdullah M Hakoun; Amir R Dehdashti; Alex B Chebl; Vikram Janardhan; Vallabh Janardhan
Journal:  Interv Neurol       Date:  2019-08-05

9.  Per-pass analysis of recanalization and good neurological outcome in thrombectomy for stroke: Systematic review and meta-analysis.

Authors:  Jorge Arturo Larco; Mehdi Abbasi; Yang Liu; Sarosh Irfan Madhani; Adnan Hussain Shahid; Ramanathan Kadirvel; Waleed Brinjikji; Luis E Savastano
Journal:  Interv Neuroradiol       Date:  2021-07-06       Impact factor: 1.764

Review 10.  Preclinical testing platforms for mechanical thrombectomy in stroke: a review on phantoms, in-vivo animal, and cadaveric models.

Authors:  Yang Liu; Mehdi Abbasi; Jorge L Arturo Larco; Ramanathan Kadirvel; David F Kallmes; Waleed Brinjikji; Luis Savastano
Journal:  J Neurointerv Surg       Date:  2021-03-15       Impact factor: 8.572