OBJECTIVE: This study evaluated the safety and efficacy of total percutaneous endovascular aortic aneurysm repair (PEVAR) with a single Perclose ProGlide device (Abbot Vascular, Santa Clara, Calif) compared with endovascular aortic repair with surgical cutdown (SEVAR).
METHODS: The study included 50 abdominal aortic aneurysm patients who were treated with PEVAR with a single Perclose ProGlide device and 96 patients treated with SEVAR. Technical success was defined as successful arterial closure of the common femoral artery without the need for adjunctive surgical or endovascular procedures. The rates of complications, including bleeding requiring transfusion, infection, pseudoaneurysm, paresthesia, and lymphocele, as well as the operating room time and hospital duration were compared between the PEVAR and SEVAR groups.
RESULTS: Technical success was obtained in all patients in the PEVAR group. One patient in the SEVAR group needed surgical repair due to access site bleeding. Complication rates were similar between the groups (4% in the PEVAR vs 8% in the SEVAR; P = .495). The PEVAR group had significantly shorter operating room times (153 ± 47 minutes vs 211 ± 88 minutes, P < .001) and hospital lengths of stay (6.7 ± 6.8 days vs 9.3 ± 4.5 days, P < .001).
CONCLUSIONS: Compared with SEVAR, PEVAR with a single ProGlide device is a safe procedure with a shorter operating room time and hospital stay, without increasing access site complications.
Copyright © 2016 Society for Vascular Surgery. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2015        PMID: 26506938     DOI: 10.1016/j.jvs.2015.08.111

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  5 in total

Review 1.  Totally percutaneous versus surgical cut-down femoral artery access for elective bifurcated abdominal endovascular aneurysm repair.

Authors:  Madelaine Gimzewska; Alexander Ir Jackson; Su Ern Yeoh; Mike Clarke
Journal:  Cochrane Database Syst Rev       Date:  2017-02-21

Review 2.  Fascia Suture Technique and Suture-mediated Closure Devices: Systematic Review.

Authors:  Georgios Karaolanis; Ioannis D Kostakis; Demetrios Moris; Viktoria-Varvara Palla; Konstantinos G Moulakakis
Journal:  Int J Angiol       Date:  2018-01-22

3.  Femoral Artery Closure Versus Surgical Cutdown for Endovascular Aortic Repair: A Single-Center Experience.

Authors:  Lin Yang; Jianlin Liu; Yanzi Li
Journal:  Med Sci Monit       Date:  2018-01-05

4.  Totally Percutaneous Access Using Perclose Proglide for Endovascular Treatment of Aortic Diseases.

Authors:  Eduardo Keller Saadi; Marina Saadi; Rodrigo Saadi; Ana Paula Tagliari; Bernardo Mastella
Journal:  Braz J Cardiovasc Surg       Date:  2017 Jan-Feb

5.  An Alternative Technique to Achieve Haemostasis During PEVAR Using Perclose ProGlide.

Authors:  Giuseppe Baldino; Federica Persi; Paolo Mortola; Amerigo Gori
Journal:  EJVES Short Rep       Date:  2018-10-30
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.