Literature DB >> 26506547

Technical note: Validation of a handheld meter for measuring β-hydroxybutyrate concentrations in plasma and serum from dairy cows.

A Pineda1, F C Cardoso2.   

Abstract

The use of handheld meters for cow-side tests for β-hydroxybutyrate (BHB) concentrations in whole blood has become common. The aim of this study was to compare serum BHB (sBHB) and plasma BHB (pBHB) concentrations analyzed using either a "gold standard" enzymatic laboratory method (LM; Randox Laboratories Ltd., Antrim, UK; cat. no. RB1007) or a handheld meter (PX; Precision Xtra, Abbott Diabetes Care Inc., Alameda, CA). Results from 374 (187 serum and 187 plasma) samples taken from Holstein cows from 11 d before (52 samples) to 5 d after parturition (137 samples) were used for the analysis. Statistical analysis was performed using the MIXED, REG, and LOGISTIC procedures of SAS (v9.4; SAS Institute Inc., Cary, NC). A linear mixed model with repeated measures was fitted for LM and PX. Regression and correlation analyses were completed to estimate the relationship and agreement between the 2 methods. Lin's concordance correlation coefficient (CCC) and Bland-Altman plots were used to evaluate agreement between LM and PX. Cross-validation by randomly splitting the data in model-building and validation sets was performed to estimate and validate the equation that predicted the LM results using PX. Receiver operating characteristic (ROC) curves were used to estimate the sensitivity (Se) and specificity (Sp) of PX at different threshold levels. The CCC was 0.74 for pBHB and 0.68 for sBHB. The 95% confidence interval of agreement of the Bland-Altman plot encompassed 96% of the difference between LM and PX for pBHB and 95% for sBHB. The mean difference for pBHB was -0.50±0.25mmol, and that for sBHB was -0.63±0.41mmol. The highest Se and Sp for PX were achieved when the threshold for ketosis was set to 1.8mmol/L for pBHB and 2.1mmol/L for sBHB. The area under the ROC curve was 0.97 for pBHB and 0.96 for sBHB. The negative bias shown by the Bland-Altman plots suggested that PX yielded higher pBHB and sBHB concentrations than the LM. However, the excellent test characteristics and area under the ROC curve close to 1 indicated that PX at the adjusted thresholds was able to accurately classify between samples <1.2mmol/L and those ≥1.2mmol/L based on the gold standard test.
Copyright © 2015 American Dairy Science Association. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  dairy cow; handheld meter; ketosis; β-hydroxybutyrate (BHB)

Mesh:

Substances:

Year:  2015        PMID: 26506547     DOI: 10.3168/jds.2015-9667

Source DB:  PubMed          Journal:  J Dairy Sci        ISSN: 0022-0302            Impact factor:   4.034


  6 in total

Review 1.  Exogenous Ketone Supplements in Athletic Contexts: Past, Present, and Future.

Authors:  Mark Evans; Tyler S McClure; Andrew P Koutnik; Brendan Egan
Journal:  Sports Med       Date:  2022-10-10       Impact factor: 11.928

2.  An Intermittent Fasting Mimicking Nutrition Bar Extends Physiologic Ketosis in Time Restricted Eating: A Randomized, Controlled, Parallel-Arm Study.

Authors:  Angie W Huang; Min Wei; Sara Caputo; Melissa L Wilson; Joseph Antoun; William C Hsu
Journal:  Nutrients       Date:  2021-04-30       Impact factor: 5.717

3.  Validation of a handheld β-hydroxybutyrate acid meter to identify hyperketonaemia in ewes.

Authors:  Carolina Akiko Sato Cabral Araújo; Antonio Humberto Hamad Minervino; Rejane Santos Sousa; Francisco Leonardo Costa Oliveira; Frederico Augusto Mazzocca Lopes Rodrigues; Clara Satsuki Mori; Enrico Lippi Ortolani
Journal:  PeerJ       Date:  2020-04-10       Impact factor: 2.984

4.  Effects of Ketosis in Dairy Cows on Blood Biochemical Parameters, Milk Yield and Composition, and Digestive Capacity.

Authors:  Wei Yang; Bingbing Zhang; Chuang Xu; Hongyou Zhang; Cheng Xia
Journal:  J Vet Res       Date:  2019-10-08       Impact factor: 1.744

5.  Reliability and diagnostic performance of a new blood ketone and glucose meter in humans.

Authors:  Andrew Ray Moore; Angelia Maleah Holland-Winkler; Jenna Kate Ansley; Eric Deiondre Hunter Boone; Megahn Kimberanne O'Reilly Schulte
Journal:  J Int Soc Sports Nutr       Date:  2021-01-07       Impact factor: 5.150

6.  Ketone Diester Ingestion Impairs Time-Trial Performance in Professional Cyclists.

Authors:  Jill J Leckey; Megan L Ross; Marc Quod; John A Hawley; Louise M Burke
Journal:  Front Physiol       Date:  2017-10-23       Impact factor: 4.566

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.