A Ravalli, L Rivas, Alfredo De la Escosura-Muñiz, J Pons, A Merkoçi, G Marrazza.
Abstract
Electrochemical aptasensors can detect different cancer biomarkers to provide point-of-care diagnosis that is low cost, rapid, specific and sensitive. In this work, we described the development of an electrochemical single-use aptasensor for detection and analysis of vascular endothelial growth factor (VEGF). Gold nanostructured graphite screen-printed electrodes were firstly modified with a mixed monolayer of a primary thiolated DNA aptamer and a spacer thiol, 6-mercapto-1-hexanol. VEGF protein was then incubated with the aptasensor. An enzyme-amplified detection scheme, based on the coupling of a streptavidin-alkaline phosphatase conjugate and secondary biotinylated aptamer was then applied. The enzyme catalyzed the hydrolysis of the electroinactive 1 -naphthyl-phosphate to 1 -naphthol; this product is electroactive and has been detected by means of differential pulse voltammetry (DPV). The aptasensor response was found to be linearly related to the target concentration between 0 and 250 nmol L(-1); the detection limit was 30 nmol L(-1). The performance of the immunoassay in terms of reproducibility and selectivity has been also studied.Entities:
Mesh:
Substances:
Year: 2015 PMID: 26504959 DOI: 10.1166/jnn.2015.10037