In vivo dosimetric impact of breast tissue expanders on post-mastectomy radiotherapy.

Harriet E Gee1,2, Fiona Bignell3, David Odgers1, Simran Gill1, Darren Martin1, Joanne Toohey1, Susan Carroll1.   

Abstract

INTRODUCTION: Temporary tissue expanders with metallic ports for gradual saline injection are increasingly employed to facilitate breast reconstruction after post-mastectomy radiotherapy (PMRT). Treatment beams therefore pass through a high-density rare-earth magnet. Measurements ex vivo suggest attenuation of dose to the skin and chest wall at clinical risk of relapse. The purpose of the study was to quantify the resulting dose reduction in vivo, compared with treatment planning system (TPS).
METHODS: Sixteen patients receiving PMRT had in vivo dosimetry prospectively performed with ethics board approval. Port was located within the expanded chest wall using the planning CT scan. Strips of radiochromic film were laid on the skin surface underneath the bolus. To aid interpretation, ex vivo measurements were also performed, including comparison with TPS predictions.
RESULTS: An average 7% reduction in dose to skin surface was measured in 15 of 16 patients. This was reproducibly located in the 'shadow' of the magnet, corresponding to each of the paths of the medial and lateral tangents. The average area was 1.07 cm(2) (range 0.39 cm(2) to 2.36 cm(2)). Ex vivo measurements confirmed attenuation of the beam in the shadow of the port. The surface area of the 'cold-spot' varied with angle of the beam relative to the metallic port. Dose attenuation in vivo differed from that predicted by the TPS.
CONCLUSION: Dose is attenuated in the 'shadow' of the tissue expander port in patients receiving PMRT. This is likely to be clinically insignificant for most, but centres should undertake appropriate measurements before utilising TPS predictions.
© 2015 The Royal Australian and New Zealand College of Radiologists.

Entities:  

Keywords:  breast neoplasm/radiotherapy; breast neoplasm/surgery; radiometry; radiotherapy dosage; tissue expansion device

Mesh:

Year:  2015        PMID: 26503758     DOI: 10.1111/1754-9485.12403

Source DB:  PubMed          Journal:  J Med Imaging Radiat Oncol        ISSN: 1754-9477            Impact factor:   1.735


  7 in total

1.  Dosimetric analysis of the effects of a temporary tissue expander on the radiotherapy technique.

Authors:  So Hyun Park; Young Suk Kim; Jinhyun Choi
Journal:  Radiol Med       Date:  2020-10-06       Impact factor: 3.469

Review 2.  Breast Implants and Radiation.

Authors:  Alexander F Mericli; Safa E Sharabi
Journal:  Semin Plast Surg       Date:  2019-10-17       Impact factor: 2.314

3.  Two-Stage Prosthetic Breast Reconstruction With Integrated Versus Remote Port Expanders: A Comparison of Complication Rates.

Authors:  Joshua H Choo; Mitchell J Buller; Michelle O'Brien; Ron Hazani; Adam Augenstein; John P Tutela; Bradon J Wilhelmi
Journal:  Eplasty       Date:  2019-04-17

4.  Magnitude and dosimetric impact of inter-fractional positional variations of the metal port of tissue expanders in postmastectomy patients treated with radiation.

Authors:  Keren Mayorov; Elsayed Ali
Journal:  Phys Imaging Radiat Oncol       Date:  2020-10-13

5.  Robustness of three external beam treatment techniques against inter-fractional positional variations of the metal port in breast tissue expanders.

Authors:  Keren Mayorov; Patricia Lacasse; Elsayed Ali
Journal:  J Appl Clin Med Phys       Date:  2021-11-22       Impact factor: 2.102

6.  Critical Factors of Dose Distribution in Breast Cancer Tomotherapy With Metallic Port Breast Tissue Expander: Image Correction, Delivery Mode, and Volume Impact.

Authors:  Hsing-Yi Lee; Yu-Hsiu Yen; Yu-Lun Tsai; Pei-Chih Tu; Chi-Ming Pu; Chia-Hong Lin; Louis Tak Lui; Suzun Shaw; Ching-Jung Wu; Hsin-Hua Nien
Journal:  Technol Cancer Res Treat       Date:  2022 Jan-Dec

7.  Modeling of the metallic port in breast tissue expanders for photon radiotherapy.

Authors:  Jihyung Yoon; Yibo Xie; David Heins; Rui Zhang
Journal:  J Appl Clin Med Phys       Date:  2018-03-30       Impact factor: 2.102

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.