Abstract

INTRODUCTION: Continuous non-invasive monitoring of cerebral blood flow (CBF) may be important during anaesthesia and several options are available. We evaluated the CerOx monitor that employs ultrasound tagged near infrared spectroscopy to estimate changes in a CBF index (CFI).
METHODS: Seven healthy males (age 21-26 years) hyperventilated and were administered phenylephrine to increase mean arterial pressure by 20-30 mmHg. Frontal lobe tissue oxygenation (ScO2) and CFI were obtained using the CerOx and mean blood flow velocity in the middle cerebral artery (MCAv mean) was determined by transcranial Doppler. Blood flow in the internal and external carotid artery (ICAf and ECAf) was determined using duplex ultrasonography and forehead skin blood flow (SkBF) and oxygenation (S skin O2) by laser Doppler and white light spectroscopy.
RESULTS: During hyperventilation MCAv mean and ICAf decreased by 44% (median; interquartile range 40-49; p = 0.016) and 46% (40-53; p = 0.03), respectively. Conversely, CFI increased by 9% (2-31; p = 0.016), while no significant change was observed in ScO2. SkBF increased by 19% (9-53; p = 0.016) and S skin O2 by 6% (1-7; p = 0.047), although ECAf was unchanged. Administration of phenylephrine was not associated with any changes in MCAv mean, ICAf, ECAf, ScO2, SkBF, S skin O2, or CFI.
CONCLUSION: The CerOx was able to detect a stable CBF during administration of phenylephrine. However, during hyperventilation MCAv mean and ICAf decreased while CFI increased, likely due to an increase in superficial tissue oxygenation. Thus, CFI does not provide an unbiased evaluation of changes in CBF.

Entities:  

Keywords:  Cerebral oximetry; external carotid artery; internal carotid artery; monitoring; skin blood flow; skin oxygen saturation

Mesh:

Substances:

Year:  2015        PMID: 26503121     DOI: 10.3109/00365513.2015.1101485

Source DB:  PubMed          Journal:  Scand J Clin Lab Invest        ISSN: 0036-5513            Impact factor:   1.713


  2 in total

1.  Assessment of a Non Invasive Brain Oximeter in Volunteers Undergoing Acute Hypoxia.

Authors:  Barry Dixon; David B MacLeod
Journal:  Med Devices (Auckl)       Date:  2020-06-30

2.  Assessment of a Non-Invasive Brain Oximeter in a Sheep Model of Acute Brain Injury.

Authors:  Barry Dixon; Renee Turner; Chris Christou
Journal:  Med Devices (Auckl)       Date:  2019-12-03
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.