[A Survey Towards Standardization of Dosimetric Verification in Intensity-modulated Radiation Therapy].

Mitsuhiro Nakamura, Hiraku Iramina, Masanori Takamiya, Tomohiro Ono, Mami Akimoto, Nobutaka Mukumoto, Yoshitomo Ishihara, Satoru Utsunomiya, Takehiro Shiinoki, Yuki Miyabe, Sayaka Sato, Hajime Monzen.   

Abstract

PURPOSE: The purpose of this study was to investigate the status of the implementation of quality assurance (QA) for intensity-modulated radiation therapy (IMRT) in Japan using a questionnaire survey.
METHODS: The questionnaire consisted of seven sections: (1) clinical uses of IMRT, (2) treatment planning systems, treatment machines, phantoms for verification and CT scanning, (3) absorbed dose verification, (4) dose distribution verification, (5) fluence map verification, (6) acceptance criteria for each verification, and (7) comments.
RESULTS: The questionnaire was completed by 129 institutions (response rate: 76.8%). IMRT was performed for prostate cancer in 125 institutions (96.9%), followed by head and neck cancer in 83 (64.3%), and brain tumors in 69 (53.5%). Although at least three individuals were engaged in IMRT QA in 77.5% of the institutions, the number of full-time persons involved in IMRT QA was one or less in 94 institutions (72.9%). This indicated that most institutions in Japan have a staff shortage. More than 90% of the institutions verified both the absorbed dose and dose distribution. The acceptance criterion for the absorbed dose verification was set to ±3% in at least 80% of the institutions. Gafchromic film was used for the majority of dose distribution verifications. The acceptance criteria for dose distribution verification mainly involved gamma analysis and a comparison of dose profiles; however, the judgment of acceptance did not depend on the results of the gamma analysis.
CONCLUSION: This survey increases our understanding of how institutions currently perform IMRT QA analysis. This understanding will help to move institutions toward more standardization of IMRT QA in Japan.

Entities:  

Mesh:

Year:  2014        PMID: 26502492

Source DB:  PubMed          Journal:  Igaku Butsuri        ISSN: 1345-5354


  2 in total

1.  Dosimetric Evaluation of the Treatment Plan on Indigenous Heterogeneous Phantoms using Analytical Anisotropic Algorithm and Acuros-XB Algorithm for Different Photon Energies.

Authors:  Vinod Kumar Gangwar; Om Prakash Gurjar; Lalit Kumar; Avinash Agarwal; Vineet Kumar Mishra; Surendra Prasad Mishra; Saket Pandey
Journal:  J Biomed Phys Eng       Date:  2022-06-01

2.  Multi-institutional evaluation using the end-to-end test for implementation of dynamic techniques of radiation therapy in Thailand.

Authors:  Lalida Tuntipumiamorn; Puangpen Tangboonduangjit; Taweap Sanghangthum; Rattapol Rangseevijitprapa; Chirasak Khamfongkhruea; Thida Niyomthai; Boonrut Vuttiprasertpong; Supranee Supanant; Nongnuch Chatchaipaiboon; Porntip Iampongpaiboon; Pitchayut Nakkrasae; Tanwiwat Jaikuna
Journal:  Rep Pract Oncol Radiother       Date:  2018-11-28