Abstract

BACKGROUND: This study compared epimysial patch electrodes with intramuscular hook electrodes using monopolar and bipolar recording configurations. The purpose was to determine which strategy transduced muscle signals with better fidelity for control of myoelectric prostheses.
METHODS: One of the two electrode styles, patch (n = 4) or hook (n = 6) was applied to the left extensor digitorum longus muscle in rats. Electrodes were evaluated at the time of placement and at monthly intervals for 4 months. Evaluations consisted of evoked electromyography signals from stimulation pulses applied to the peroneal and tibial nerves in both monopolar and bipolar recording configurations.
RESULTS: Compared with hook electrodes, patch electrodes recorded larger signals of interest and minimized muscle tissue injury. A bipolar electrode configuration significantly reduced signal noise when compared with a monopolar configuration.
CONCLUSION: Epimysial patch electrodes outperform intramuscular hook electrodes during chronic skeletal muscle implantation. Thieme Medical Publishers 333 Seventh Avenue, New York, NY 10001, USA.

Entities:  

Mesh:

Year:  2015        PMID: 26502083     DOI: 10.1055/s-0035-1565248

Source DB:  PubMed          Journal:  J Reconstr Microsurg        ISSN: 0743-684X            Impact factor:   2.873


  5 in total

Review 1.  The Evolution of Neuroprosthetic Interfaces.

Authors:  Dayo O Adewole; Mijail D Serruya; James P Harris; Justin C Burrell; Dmitriy Petrov; H Isaac Chen; John A Wolf; D Kacy Cullen
Journal:  Crit Rev Biomed Eng       Date:  2016

Review 2.  Selectivity and Longevity of Peripheral-Nerve and Machine Interfaces: A Review.

Authors:  Usman Ghafoor; Sohee Kim; Keum-Shik Hong
Journal:  Front Neurorobot       Date:  2017-10-31       Impact factor: 2.650

3.  Surgical Algorithm for Neuroma Management: A Changing Treatment Paradigm.

Authors:  Kyle R Eberlin; Ivica Ducic
Journal:  Plast Reconstr Surg Glob Open       Date:  2018-10-16

Review 4.  Cut wires: The Electrophysiology of Regenerated Tissue.

Authors:  Alexis L Lowe; Nitish V Thakor
Journal:  Bioelectron Med       Date:  2021-02-23

5.  Adjacent regenerative peripheral nerve interfaces produce phase-antagonist signals during voluntary walking in rats.

Authors:  Daniel Ursu; Andrej Nedic; Melanie Urbanchek; Paul Cederna; R Brent Gillespie
Journal:  J Neuroeng Rehabil       Date:  2017-04-24       Impact factor: 4.262

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.