Scintillation properties of Ca co-doped L(Y)SO:Ce between 193 K and 373 K for TOF-PET/MRI.

David N Ter Weele1, Dennis R Schaart1, Pieter Dorenbos1.   

Abstract

Entities:  

Year:  2014        PMID: 26501595      PMCID: PMC4545838          DOI: 10.1186/2197-7364-1-S1-A10

Source DB:  PubMed          Journal:  EJNMMI Phys        ISSN: 2197-7364


× No keyword cloud information.
Time-of-flight Positron Emission Tomography (TOF-PET) and TOF-PET/MRI require scintillators with high light yield, short decay time, and short rise time in order to obtain high timing resolution. LSO:Ce and LYSO:Ce are commonly used. Ca co-doped LSO:Ce shows improved scintillation properties. The decay time constant of LSO:Ce,0.2%Ca (~33 ns) is shorter than standard LSO:Ce (~38-40 ns), and it has about 15% higher light yield. We measured scintillation pulse shapes and photoelectron yields of LSO:Ce, LSO:Ce,0.2%Ca, LYSO:Ce, LYSO:Ce,20ppmCa, LYSO:0.11%Ce,0.2%Mg, and LYSO:0.2%Ce,0.2%Ca at temperatures ranging from 193 K to 373 K. To study rise times we built a set-up in which samples are excited by 100 ps (FWHM) x-ray pulses. Figure 1 shows that the scintillation rise time of LSO:Ce,0.2%Ca is constant between 273 K and 373 K, while the decay time constant decreases. Figure 2 shows that the light yield of LSO:Ce,0.2%Ca decreases with rising temperature. We noticed that the decrease in light yield is entirely caused by non-radiative decay of excited Ce3+ centres. This quenching mechanism does not worsen the timing resolution, as opposed to recombination of electron-hole pairs. It follows that the timing resolution of LSO:Ce(,0.2%Ca) is constant over the entire temperature range, in accordance with the timing model described in [1].
Figure 1

Scintillation pulse shape of LSO:Ce,0.2%Ca.

Figure 2

Normalized scintillation properties as a function of temperature.

Scintillation pulse shape of LSO:Ce,0.2%Ca. Normalized scintillation properties as a function of temperature.
  1 in total

1.  The lower bound on the timing resolution of scintillation detectors.

Authors:  Stefan Seifert; Herman T van Dam; Dennis R Schaart
Journal:  Phys Med Biol       Date:  2012-03-13       Impact factor: 3.609

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.