Literature DB >> 26497254

Multi-layer Flow-modulating Stents for Thoraco-abdominal and Peri-renal Aneurysms: The UK Pilot Study.

C Lowe1, A Worthington2, F Serracino-Inglott3, R Ashleigh4, C McCollum5.   

Abstract

OBJECTIVE: There remains a population of patients with aortic aneurysms that cannot be treated by conventional endovascular means. Multi-layer flow modulating stents (MFMS) are a novel approach for the treatment of aortic aneurysm; this study reports outcomes of a UK pilot study of first-generation MFMS in thoraco-abdominal (TAAA) and perirenal aneurysms (PAA) in patients who were also unfit for open surgery.
METHODS: Patients with TAAA and PAA unfit for open surgery and with no conventional options for endovascular repair were recruited. Follow-up included CTA at 1, 3, 6, and 12 months, then annually. Outcome measures included 30 day mortality, growth-free survival, branch vessel patency, complications, re-intervention, and maximal aortic diameter.
RESULTS: MFMS were implanted in 14 patients (6 PAA, 8 TAAA) between October 2011 and March 2014 with one (7%) 30 day death and 11 (79%) surviving to 12 months. The median aneurysm growth was 9 mm in the first 12 months following implantation. On mean follow-up of 22.8 months, seven (50%) patients had died including one confirmed rupture. AAA diameter remained stable in only two of the surviving patients. Fifty of 51 covered aortic branches remained patent with no embolic episodes or symptoms of ischaemia in any patient. MFMS dislocation occurred in four patients, leading to re-intervention in two. A total of six re-interventions were performed in five patients (35%) with one post-re-intervention death.
CONCLUSION: These first-generation MFMS were unstable and dislocated frequently. It is uncertain whether MFMS implantation influenced the natural history of these aneurysms as none decreased in size, but two remain stable after a mean of 22.8 months. Although side branch patency was maintained, our results do not support the continued use of these first-generation devices. Further development is needed if this technology is to have a role in treatment of aortic aneurysm.
Copyright © 2015 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Aortic aneurysm; Endovascular repair; Multi-layered stent

Mesh:

Year:  2015        PMID: 26497254     DOI: 10.1016/j.ejvs.2015.09.014

Source DB:  PubMed          Journal:  Eur J Vasc Endovasc Surg        ISSN: 1078-5884            Impact factor:   7.069


  6 in total

1.  Experimental Analysis of Pressure and Flow Alterations During and After Insertion of a Multilayer Flow Modulator into an AAA Model with Incorporated Branch.

Authors:  Simon Tupin; Kei Takase; Makoto Ohta
Journal:  Cardiovasc Intervent Radiol       Date:  2021-04-27       Impact factor: 2.740

2.  Multilayer stents affect the final diameter of aortic aneurysms and maintain renal artery patency for a short time in a swine experimental model.

Authors:  Anna Paula Weinhardt Baptista-Strazzi; Ricardo Aun; Igor Rafael Sincos; Allana M Tobita; Maria Fernanda Cassino Portugal; Vitória Penido de Paula; Oskar Kaufmann; Nelson Wolosker
Journal:  Clinics (Sao Paulo)       Date:  2021-05-21       Impact factor: 2.365

3.  Hemodynamics study of a multilayer stent for the treatment of aneurysms.

Authors:  Yan Xiong; Xuhong Wang; Wentao Jiang; Xiaobao Tian; Qingyuan Wang; Yubo Fan; Yu Chen
Journal:  Biomed Eng Online       Date:  2016-12-28       Impact factor: 2.819

4.  Is endovascular treatment with multilayer flow modulator stent insertion a safe alternative to open surgery for high-risk patients with thoracoabdominal aortic aneurysm?

Authors:  Carolline Pinto; George Garas; Leanne Harling; Ara Darzi; Roberto Casula; Thanos Athanasiou
Journal:  Ann Med Surg (Lond)       Date:  2017-01-31

5.  Endovascular management of renal artery aneurysms using the multilayer flow modulator.

Authors:  Sherif Sultan; Mahmoud Basuoniy Alawy; Rita Flaherty; Edel P Kavanagh; Mohamed Elsherif; Ala Elhelali; Florian Stefanov; Violet Lundon; Niamh Hynes
Journal:  Open Heart       Date:  2016-03-10

6.  One-Year Single-Center Results of the Multilayer Flow Modulator Stents for the Treatment of Type B Aortic Dissection.

Authors:  Victor S Costache; Jorn P Meekel; Andreea Costache; Tatiana Melnic; Cristian Bucurenciu; Anca Chitic; Gabriela Candea; Crina Solomon; Kak K Yeung
Journal:  J Endovasc Ther       Date:  2020-09-01       Impact factor: 3.487