Transcutaneous bilirubinometry is a reliably conservative method of assessing neonatal jaundice.

Ganga L Srinivas1, Christina Dibattista Cuff1, Myla D Ebeling1, James T Mcelligott1.   

Abstract

OBJECTIVE: In order to reduce invasive testing in newborns prior to discharge, we tested the direction of the correlation between transcutaneous bilirubin (TcB) and total serum bilirubin (TSB), the likelihood of missing high TSBs with a raised threshold for confirmatory testing, and also calculated potential cost savings from fewer laboratory testing.
METHODS: We performed a cross-sectional analysis of single paired TcB and TSB results measured at 36 ± 2 h of life in neonates ≥37 weeks admitted only to the Level 1 nursery. TcB was measured using the BiliChek® meter.
RESULTS: Of the 552 infants, 512 (92.8%) had TSB levels below TcB values. Correlation between TcB and TSB was 0.69. If TSB confirmation was to be performed at 11.7 mg/dL (medium risk threshold for phototherapy), the negative predictive value was 99.4%, with a potential cost savings of $6555.00 ($1500.00 per 100 patients). Of the 495 infants with TcB <11.7 mg/dL, only 3 had TSB levels higher than 11.7 mg/dL, and none met phototherapy threshold for low risk infants.
CONCLUSIONS: TcB screening at our institution has a high negative predictive value, and can be used as a stand-alone test until values are close to phototherapy threshold, thus reducing invasive testing and cost.

Entities:  

Keywords:  BiliChek® meter; correlation; hyperbilirubinemia; newborn

Mesh:

Substances:

Year:  2015        PMID: 26483074     DOI: 10.3109/14767058.2015.1095881

Source DB:  PubMed          Journal:  J Matern Fetal Neonatal Med        ISSN: 1476-4954


  2 in total

1.  Cost savings with transcutaneous screening versus total serum bilirubin measurement for newborn jaundice in hospital and community settings: a cost-minimization analysis.

Authors:  Stephanie McClean; Krista Baerg; Julie Smith-Fehr; Michael Szafron
Journal:  CMAJ Open       Date:  2018-07-27

2.  Transcutaneous bilirubin measurement in healthy Saudi term newborns.

Authors:  Saad A Alsaedi
Journal:  Saudi Med J       Date:  2016-02       Impact factor: 1.484

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.