Literature DB >> 26482816

ACR White Paper-Based Comprehensive Dose Reduction Initiative Is Associated With a Reversal of the Upward Trend in Radiation Dose for Chest CT.

Udit Rawat1, Stuart L Cohen2, Jeffrey M Levsky1, Linda B Haramati1.   

Abstract

PURPOSE: In 2010, the authors' department implemented a comprehensive dose reduction strategy based on the ACR white paper on radiation dose in medicine. The aim of this study was to evaluate the effectiveness of the dose reduction program.
METHODS: In total, 1,234 adult chest CT scans from 2007 to 2012 were analyzed retrospectively, with institutional review board approval and a waiver of the requirement for informed consent. The primary outcome was effective dose in millisieverts during the three-year periods before (2007-2009) and after (2010-2012) dose reduction implementation. Dose trends were analyzed by fitted linear modeling. The use and effects on total exposure of dose reduction strategies (high pitch, adaptive statistical iterative reconstruction [ASIR], and low tube voltage) were analyzed.
RESULTS: The overall mean dose for chest CT was 7.3 ± 5.1 mSv. The mean dose decreased by 30%, from 9.2 mSv (2007-2009) to 6.5 mSv (2010-2012) (P < .001). From 2007 to 2009, the mean dose increased by 1.2 mSv per year (P < .01). From 2010 to 2012, the mean dose decreased by 1.1 mSv per year (P < 0.01). High-pitch technique, ASIR, and low tube voltage increased significantly after dose reduction implementation. High pitch and ASIR were significantly associated with a reduced dose, whereas the effect of reduced voltage was not significant.
CONCLUSIONS: Reductions in radiation exposure from medical imaging rely on ongoing technical developments and consistent, vigilant use of dose reduction strategies. This comprehensive dose reduction strategy significantly reduced radiation exposure from chest CT. Annual increases in radiation dose reversed after the strategy was implemented and continued to decline over the study period.
Copyright © 2015 American College of Radiology. Published by Elsevier Inc. All rights reserved.

Keywords:  Dose reduction; chest CT; radiation dose

Mesh:

Year:  2015        PMID: 26482816     DOI: 10.1016/j.jacr.2015.07.022

Source DB:  PubMed          Journal:  J Am Coll Radiol        ISSN: 1546-1440            Impact factor:   5.532


  2 in total

1.  Optimizing Radiation Doses for Computed Tomography Across Institutions: Dose Auditing and Best Practices.

Authors:  Joshua Demb; Philip Chu; Thomas Nelson; David Hall; Anthony Seibert; Ramit Lamba; John Boone; Mayil Krishnam; Christopher Cagnon; Maryam Bostani; Robert Gould; Diana Miglioretti; Rebecca Smith-Bindman
Journal:  JAMA Intern Med       Date:  2017-06-01       Impact factor: 21.873

2.  Proposal of a low-dose, long-pitch, dual-source chest CT protocol on third-generation dual-source CT using a tin filter for spectral shaping at 100 kVp for CoronaVirus Disease 2019 (COVID-19) patients: a feasibility study.

Authors:  Andrea Agostini; Chiara Floridi; Alessandra Borgheresi; Myriam Badaloni; Paolo Esposto Pirani; Filippo Terilli; Letizia Ottaviani; Andrea Giovagnoni
Journal:  Radiol Med       Date:  2020-04-01       Impact factor: 6.313

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.