Gas transfer model to design a ventilator for neonatal total liquid ventilation.

Mirko Bonfanti1, Antonio Cammi2, Paola Bagnoli3.   

Abstract

The study was aimed to optimize the gas transfer in an innovative ventilator for neonatal Total Liquid Ventilation (TLV) that integrates the pumping and oxygenation functions in a non-volumetric pulsatile device made of parallel flat silicone membranes. A computational approach was adopted to evaluate oxygen (O2) and carbon dioxide (CO2) exchanges between the liquid perfluorocarbon (PFC) and the oxygenating gas, as a function of the geometrical parameter of the device. A 2D semi-empirical model was implemented to this purpose using Comsol Multiphysics to study both the fluid dynamics and the gas exchange in the ventilator. Experimental gas exchanges measured with a preliminary prototype were compared to the simulation outcomes to prove the model reliability. Different device configurations were modeled to identify the optimal design able to guarantee the desired gas transfer. Good agreement between experimental and simulation outcomes was obtained, validating the model. The optimal configuration, able to achieve the desired gas exchange (ΔpCO2 = 16.5 mmHg and ΔpO2 = 69 mmHg), is a device comprising 40 modules, 300 mm in length (total exchange area = 2.28 m(2)). With this configuration gas transfer performance is satisfactory for all the simulated settings, proving good adaptability of the device.
Copyright © 2015 IPEM. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Gas exchange; Neonatal ventilator; Numerical model; Perfluorocarbon; Total liquid ventilation

Mesh:

Substances:

Year:  2015        PMID: 26475493     DOI: 10.1016/j.medengphy.2015.09.003

Source DB:  PubMed          Journal:  Med Eng Phys        ISSN: 1350-4533            Impact factor:   2.242


  1 in total

1.  The Use of Stepper Motor-Controlled Proportional Valve for Fio2 Calculation in the Ventilator and its Control with Fuzzy Logic.

Authors:  Adem Gölcük; İnan Güler
Journal:  J Med Syst       Date:  2016-11-05       Impact factor: 4.460

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.