Literature DB >> 26472181

Evaluation of the Double-Lumen Tube Vivasight-DL (DLT-ETView): A Prospective Single-Center Study.

Julien Massot1, Virginie Dumand-Nizard2, Marc Fischler3, Morgan Le Guen2.   

Abstract

OBJECTIVES: VivaSight-DL (DLT-ETView) is a single-use double-lumen tube (DLT) with an integrated camera visualizing the carina continuously, which could reduce the need for a routine fiberoptic bronchoscopy. The objective of this study was to evaluate its rate of correct positioning.
DESIGN: A prospective, observational study.
SETTING: A university hospital. PARTICIPANTS: 170 patients undergoing a thoracic surgical procedure.
INTERVENTIONS: Selective left bronchus intubation.
MEASUREMENTS AND MAIN RESULTS: The main outcome measure was a double criteria assessed by fiberoptic bronchoscopy: the bronchial cuff in the left main bronchus and end of the bronchial lumen upstream of the lobar bifurcation. Between September 2013 and April 2014, 84 patients of the planned 170 were included. Seven patients were excluded: 3 for failed intubation, 3 for protocol violation, and 1 because the DLT melted before insertion. The study was terminated after this event. Seventy-six patients (99%) had correct positioning, with a median margin of safety of 20 mm (interquartile range: 15-27) in the surgical position. In 40 patients (53%), malpositioning required mobilization of the tube at least once intraoperatively.
CONCLUSIONS: The tube was well positioned in almost all patients. Continuous visualization of the carina is a major improvement for patient care as intraoperative displacement can be diagnosed immediately and corrected. However, an incident induced premature ending of the study.
Copyright © 2015 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  VivaSight; bronchoscopy; double-lumen tube; intratracheal; intubation; one-lung ventilation

Mesh:

Year:  2015        PMID: 26472181     DOI: 10.1053/j.jvca.2015.04.031

Source DB:  PubMed          Journal:  J Cardiothorac Vasc Anesth        ISSN: 1053-0770            Impact factor:   2.628


  9 in total

Review 1.  Development of endotracheal intubation devices for patients with tumors.

Authors:  Yang Liu; Yang Zhang; Bin Zhu; Wenyun Xu; Yi Yang; Zui Zou
Journal:  Am J Cancer Res       Date:  2022-06-15       Impact factor: 5.942

2.  Recommendations from the Italian intersociety consensus on Perioperative Anesthesa Care in Thoracic surgery (PACTS) part 2: intraoperative and postoperative care.

Authors:  Federico Piccioni; Andrea Droghetti; Alessandro Bertani; Cecilia Coccia; Antonio Corcione; Angelo Guido Corsico; Roberto Crisci; Carlo Curcio; Carlo Del Naja; Paolo Feltracco; Diego Fontana; Alessandro Gonfiotti; Camillo Lopez; Domenico Massullo; Mario Nosotti; Riccardo Ragazzi; Marco Rispoli; Stefano Romagnoli; Raffaele Scala; Luigia Scudeller; Marco Taurchini; Silvia Tognella; Marzia Umari; Franco Valenza; Flavia Petrini
Journal:  Perioper Med (Lond)       Date:  2020-10-23

3.  McGrath Series 5 videolaryngoscope vs Airtraq DL videolaryngoscope for double-lumen tube intubation: A randomized trial.

Authors:  Li Wan; Mingfeng Liao; Li Li; Wei Qian; Rong Hu; Kun Chen; Chuanhan Zhang; Wenlong Yao
Journal:  Medicine (Baltimore)       Date:  2016-12       Impact factor: 1.889

4.  The use of ETView endotracheal tube for surveillance after tube positioning in patients undergoing lobectomy, randomized trial.

Authors:  Hui-Hui Liu; Fang Dong; Jia-Yi Liu; Jian-Qi Wei; Yan-Kui Huang; Yong Wang; Tao Zhou; Wu-Hua Ma
Journal:  Medicine (Baltimore)       Date:  2018-12       Impact factor: 1.817

5.  A Cost-Effectiveness Analysis Comparing the VivaSight Double-Lumen Tube and a Conventional Double-Lumen Tube in Adult Patients Undergoing Thoracic Surgery Involving One-Lung Ventilation.

Authors:  Sara Larsen; Jimmy Højberg Holm; Tove Nørgaard Sauer; Claus Andersen
Journal:  Pharmacoecon Open       Date:  2020-03

6.  Comparing the rate of fiberoptic bronchoscopy use with a video double lumen tube versus a conventional double lumen tube-a randomized controlled trial.

Authors:  Akinjide Onifade; Dlorean Lemon-Riggs; Aaron Smith; Taylor Pak; Jessica Pruszynski; Scott Reznik; Tiffany S Moon
Journal:  J Thorac Dis       Date:  2020-11       Impact factor: 2.895

Review 7.  Optical Devices in Tracheal Intubation-State of the Art in 2020.

Authors:  Jan Matek; Frantisek Kolek; Olga Klementova; Pavel Michalek; Tomas Vymazal
Journal:  Diagnostics (Basel)       Date:  2021-03-22

8.  Intubation with vivasight double-lumen tube versus conventional double-lumen tube in adult patients undergoing lung resection: A retrospective analysis.

Authors:  Manuel Granell; Giulia Petrini; Pablo Kot; Mercedes Murcia; Javier Morales; Ricardo Guijarro; José A de Andrés
Journal:  Ann Card Anaesth       Date:  2022 Jul-Sep

Review 9.  [Application and Research Progress of Video Double-lumen Tube in Thoracic Surgery].

Authors:  Cheng Shen; Peng Liang; Guowei Che
Journal:  Zhongguo Fei Ai Za Zhi       Date:  2022-08-20
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.