Literature DB >> 26471141

In Vitro Testing of an Implantable Wireless Telemetry System for Long-Term Electromyography Recordings in Large Animals.

Lukas Kneisz1,2, Ewald Unger1, Hermann Lanmüller1, Winfried Mayr1.   

Abstract

Multichannel bio-signal recording in undisturbed in vivo conditions is a frequent demand in experimental work for development of methodology and associated equipment for functional electrical stimulation (FES) application, limb prosthesis, and diagnostic tools in contemporary rehabilitation efforts. Intramuscular electromyogram (EMG) recordings can provide comprehensive insight in complex interactions of agonistic and antagonistic muscles during movement tasks and in contrast act as reliable control signals for both neuroprosthesis and mechanical prosthesis. We fabricated a fully implantable device, which is capable of recording electromyography signals from inside a body and transmit these signals wirelessly to an external receiver. The developed analog front end uses only two electrodes per channel, provides a gain of 60 dB, and incorporates a band pass filter with lower cut-off frequency of 4 Hz and upper cut-off frequency of 480 Hz. The bidirectional wireless data link, which operates in the 2.4 GHz Industrial, Scientific and Medical band, is designed for transmission distances of 10 m using an application data rate of 1 kSps for each of the two channels. Performed in vitro tests with the devices coated in epoxy resin and inserted into a phantom with tissue-equivalent characteristics confirmed the functionality of our concept and the measurement results are consistent with those from preceding simulations.
Copyright © 2015 International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

Keywords:  Biomedical telemetry; Intramuscular electromyography; Received signal strength indicator

Mesh:

Year:  2015        PMID: 26471141     DOI: 10.1111/aor.12626

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  2 in total

1.  An Implantable System For Chronic In Vivo Electromyography.

Authors:  David Zealear; Yike Li; Shan Huang
Journal:  J Vis Exp       Date:  2020-04-21       Impact factor: 1.355

2.  Investigation into pathophysiology of naturally occurring palatal instability and intermittent dorsal displacement of the soft palate (DDSP) in racehorses: Thyro-hyoid muscles fatigue during exercise.

Authors:  Marta Cercone; Emil Olsen; Justin D Perkins; Jonathan Cheetham; Lisa M Mitchell; Norm G Ducharme
Journal:  PLoS One       Date:  2019-10-25       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.