Literature DB >> 26471041

Condensed Phase Membrane Introduction Mass Spectrometry with Direct Electron Ionization: On-line Measurement of PAHs in Complex Aqueous Samples.

Veronica Termopoli1, Giorgio Famiglini, Pierangela Palma, Achille Cappiello, Gregory W Vandergrift, Erik T Krogh, Chris G Gill.   

Abstract

Polycyclic aromatic hydrocarbons (PAHs) are USEPA regulated priority pollutants. Their low aqueous solubility requires very sensitive analytical methods for their detection, typically involving preconcentration steps. Presented is the first demonstrated ‘proof of concept’ use of condensed phase membrane introduction mass spectrometry (CP-MIMS) coupled with direct liquid electron ionization (DEI) for the direct, on-line measurement of PAHs in aqueous samples. DEI is very well suited for the ionization of PAHs and other nonpolar compounds, and is not significantly influenced by the co-elution of matrix components. Linear calibration data for low ppb levels of aqueous naphthalene, anthracene, and pyrene is demonstrated, with measured detection limits of 4 ppb. Analytical response times (t10%–90% signal rise) ranged from 2.8 min for naphthalene to 4.7 min for pyrene. Both intra- and interday reproducibility has been assessed (<3% and 5% RSD, respectively). Direct measurements of ppb level PAHs spiked in a variety of real, complex environmental sample matrices is examined, including natural waters, sea waters, and a hydrocarbon extraction production waste water sample. For these spiked, complex samples, direct PAH measurement by CP-MIMS-DEI yielded minimal signal suppression from sample matrix effects (81%–104%). We demonstrate the use of this analytical approach to directly monitor real-time changes in aqueous PAH concentrations with potential applications for continuous on-line monitoring strategies and binding/adsorption studies in heterogeneous samples.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 26471041     DOI: 10.1007/s13361-015-1285-9

Source DB:  PubMed          Journal:  J Am Soc Mass Spectrom        ISSN: 1044-0305            Impact factor:   3.109


  26 in total

Review 1.  Membrane introduction mass spectrometry: trends and applications.

Authors:  R C Johnson; R G Cooks; T M Allen; M E Cisper; P H Hemberger
Journal:  Mass Spectrom Rev       Date:  2000 Jan-Feb       Impact factor: 10.946

Review 2.  Sorptive sample preparation -- a review.

Authors:  E Baltussen; C A Cramers; P J F Sandra
Journal:  Anal Bioanal Chem       Date:  2002-04-09       Impact factor: 4.142

3.  Liquid chromatography mass spectrometry with supersonic molecular beams

Authors: 
Journal:  J Am Soc Mass Spectrom       Date:  2000-06       Impact factor: 3.109

4.  An electrospray membrane probe for the analysis of volatile and semi-volatile organic compounds in water.

Authors:  Emily VanHassel; Mark E Bier
Journal:  Rapid Commun Mass Spectrom       Date:  2007       Impact factor: 2.419

5.  Easy ambient sonic-spray ionization-membrane interface mass spectrometry for direct analysis of solution constituents.

Authors:  Renato Haddad; Regina Sparrapan; Tapio Kotiaho; Marcos N Eberlin
Journal:  Anal Chem       Date:  2008-01-08       Impact factor: 6.986

6.  Handheld miniature ion trap mass spectrometers.

Authors:  Zheng Ouyang; Robert J Noll; R Graham Cooks
Journal:  Anal Chem       Date:  2009-04-01       Impact factor: 6.986

7.  Advanced liquid chromatography-mass spectrometry interface based on electron ionization.

Authors:  A Cappiello; G Famiglini; E Pierini; P Palma; H Trufelli
Journal:  Anal Chem       Date:  2007-06-15       Impact factor: 6.986

8.  Characterization of a condensed-phase membrane introduction mass spectrometry (CP-MIMS) interface using a methanol acceptor phase coupled with electrospray ionization for the continuous on-line quantitation of polar, low-volatility analytes at trace levels in complex aqueous samples.

Authors:  K D Duncan; E P B McCauley; E T Krogh; C G Gill
Journal:  Rapid Commun Mass Spectrom       Date:  2011-05-15       Impact factor: 2.419

9.  A coaxially heated membrane introduction mass spectrometry interface for the rapid and sensitive on-line measurement of volatile and semi-volatile organic contaminants in air and water at parts-per-trillion levels.

Authors:  Alexander J Thompson; A Skye Creba; Robyn M Ferguson; Erik T Krogh; Christopher G Gill
Journal:  Rapid Commun Mass Spectrom       Date:  2006       Impact factor: 2.419

10.  Laser desorption-membrane introduction mass spectrometry.

Authors:  M H Soni; J H Callahan; S W McElvany
Journal:  Anal Chem       Date:  1998-08-01       Impact factor: 6.986