A candidate reference method using ICP-MS for sweat chloride quantification.

Jake T Collie, R John Massie, Oliver A H Jones, Paul D Morrison, Ronda F Greaves.   

Abstract

BACKGROUND: The aim of the study was to develop a method for sweat chloride (Cl) quantification using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) to present to the Joint Committee for Traceability in Laboratory Medicine (JCTLM) as a candidate reference method for the diagnosis of cystic fibrosis (CF).
METHODS: Calibration standards were prepared from sodium chloride (NaCl) to cover the expected range of sweat Cl values. Germanium (Ge) and scandium (Sc) were selected as on-line (instrument based) internal standards (IS) and gallium (Ga) as the off-line (sample based) IS. The method was validated through linearity, accuracy and imprecision studies as well as enrolment into the Royal College of Pathologists of Australasia Quality Assurance Program (RCPAQAP) for sweat electrolyte testing.
RESULTS: Two variations of the ICP-MS method were developed, an on-line and off-line IS, and compared. Linearity was determined up to 225 mmol/L with a limit of quantitation of 7.4 mmol/L. The off-line IS demonstrated increased accuracy through the RCPAQAP performance assessment (CV of 1.9%, bias of 1.5 mmol/L) in comparison to the on-line IS (CV of 8.0%, bias of 3.8 mmol/L). Paired t-tests confirmed no significant differences between sample means of the two IS methods (p=0.53) or from each method against the RCPAQAP target values (p=0.08 and p=0.29).
CONCLUSIONS: Both on and off-line IS methods generated highly reproducible results and excellent linear comparison to the RCPAQAP target results. ICP-MS is a highly accurate method with a low limit of quantitation for sweat Cl analysis and should be recognised as a candidate reference method for the monitoring and diagnosis of CF. Laboratories that currently practice sweat Cl analysis using ICP-MS should include an off-line IS to help negate any pre-analytical errors.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 26457780     DOI: 10.1515/cclm-2015-0506

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  5 in total

1.  Australasian Guideline (2nd Edition): an Annex to the CLSI and UK Guidelines for the Performance of the Sweat Test for the Diagnosis of Cystic Fibrosis.

Authors:  John Massie; Ronda Greaves; Michael Metz; Veronica Wiley; Peter Graham; Samantha Shepherd; Richard Mackay
Journal:  Clin Biochem Rev       Date:  2017-11

2.  Sweat chloride quantification using capillary electrophoresis.

Authors:  Patricia Dubot; Jing Liang; Jacobé Dubs; Yohann Missiak; Cédric Sarazin; François Couderc; Elizabeth Caussé
Journal:  Pract Lab Med       Date:  2018-12-07

3.  Ion chromatography as candidate reference method for the determination of chloride in human serum.

Authors:  Min Shen; Minmin Tu; Wei Zhang; Jihua Zou; Man Zhang; Zheng Cao; Bingde Zou
Journal:  J Clin Lab Anal       Date:  2020-07-24       Impact factor: 2.352

4.  Physicochemical and Colligative Investigation of α (Shrimp Shell)- and β (Squid Pen)-Chitosan Membranes: Concentration-Gradient-Driven Water Flux and Ion Transport for Salinity Gradient Power and Separation Process Operations.

Authors:  Clifford R Merz
Journal:  ACS Omega       Date:  2019-12-05

5.  Sweat chloride assay by inductively coupled plasma mass spectrometry: a confirmation test for cystic fibrosis diagnosis.

Authors:  Antonella Marvelli; Beatrice Campi; Gianfranco Mergni; Maria Elisa Di Cicco; Paola Turini; Paolo Scardina; Riccardo Zucchi; Massimo Pifferi; Giovanni Taccetti; Aldo Paolicchi; Giancarlo la Marca; Alessandro Saba
Journal:  Anal Bioanal Chem       Date:  2020-07-21       Impact factor: 4.142