Analytical and clinical performance of the new Fujirebio 25-OH vitamin D assay, a comparison with liquid chromatography-tandem mass spectrometry (LC-MS/MS) and three other automated assays.

Lanja Saleh, Daniel Mueller, Arnold von Eckardstein.   

Abstract

BACKGROUND: We evaluated the analytical and clinical performance of the new Lumipulse® G 25-OH vitamin D assay from Fujirebio, and compared it to a liquid chromatography-tandem mass spectrometry (LC-MS/MS) method and three other commercial automated assays.
METHODS: Total 25 hydroxy vitamin D (25(OH)D) levels were measured in 100 selected serum samples from our routine analysis with Fujirebio 25(OH)D assay. The results were compared with those obtained with LC-MS/MS and three other automated 25(OH)D assays (Abbott, Beckman, and Roche). The accuracy of each assay tested was evaluated against a Labquality reference serum panel for 25(OH)D (Ref!25OHD; University of Ghent).
RESULTS: Intra- and inter-day imprecision of the Fujirebio 25(OH)D assay was <5%. Fujirebio 25(OH)D assay showed the highest correlation among the assays tested with the LC-MS/MS method (R=0.986). The mean relative bias obtained was -15.6% (Fujirebio), -12.7% (Beckman), -2.1% (Abbott) and 9.7% (Roche) as compared to LC-MS/MS. Comparison with the Labquality certified reference serum panel yielded a mean bias of -11.8% (Fujirebio), -14.1% (Beckman), 4.4% (Abbott) and 3.2% (Roche), respectively. Compared to LC-MS/MS, the sensitivity of different methods in detecting vitamin D insufficiency (<50 nmol/L) varied from 100% for the Fujirebio assay to 72.7% for Roche, and specificity ranged from 94.4% for Roche to 87.6% for Beckman.
CONCLUSIONS: The Lumipulse G 25-OH vitamin D assay from Fujirebio demonstrated a good correlation with LC-MS/MS and some immunoassays. The performance of the assay is well-suited for routine 25(OH)D measurement in clinical serum samples. A correction for the observed negative bias vs. LC-MS/MS could be considered.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 26457778     DOI: 10.1515/cclm-2015-0427

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  4 in total

1.  Evaluation of the new restandardized Abbott Architect 25-OH Vitamin D assay in vitamin D-insufficient and vitamin D-supplemented individuals.

Authors:  Wijtske Annema; Albina Nowak; Arnold von Eckardstein; Lanja Saleh
Journal:  J Clin Lab Anal       Date:  2017-09-19       Impact factor: 2.352

2.  Association of circulating 25-Hydroxyvitamin D and its related genetic variations with hepatocellular carcinoma incidence and survival.

Authors:  Hui Liu; Xia Jiang; Qiaohua Qiao; Liying Chen; Koichi Matsuda; Guangyi Jiang; Tunan Yu; Yifan Wang; Hui Lin; Xiao Liang; Xiujun Cai
Journal:  Ann Transl Med       Date:  2020-09

3.  Comparative study of two immunoassays used for the determination of serum vitamin D.

Authors:  Raquel Lahoz; Juan Pelegrín Sánchez; Silvia Górriz; Pilar Calmarza
Journal:  Pract Lab Med       Date:  2021-06-23

4.  Consensus statement from 2nd International Conference on Controversies in Vitamin D.

Authors:  A Giustina; R A Adler; N Binkley; J Bollerslev; R Bouillon; B Dawson-Hughes; P R Ebeling; D Feldman; A M Formenti; M Lazaretti-Castro; C Marcocci; R Rizzoli; C T Sempos; J P Bilezikian
Journal:  Rev Endocr Metab Disord       Date:  2020-03       Impact factor: 6.514

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.