Abstract
We evaluated Abbott's RealTime assay for the detection of Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (GC) in the urethra, oropharynx, and rectum of 260 men who have sex with men. Compared with Hologic's AC2, RealTime had good agreement for detecting CT and GC. Overall, there were 25 CT and 44 GC AC2 positives, and 26 CT and 38 GC RealTime positives. For total negatives, there were 742 CT and 725 GC for AC2, 744 CT and 724 GC for RealTime.Entities:
Mesh:
Substances:
Year: 2015 PMID: 26457488 DOI: 10.1097/OLQ.0000000000000360