Literature DB >> 26456253

Modified Haigis Formula Effective Lens Position Equation for Ciliary Sulcus-Implanted Intraocular Lenses.

Youngsub Eom1, Jong Suk Song1, Hyo Myung Kim2.   

Abstract

PURPOSE: To investigate the effect of a modified Haigis formula effective lens position equation on improving refractive outcomes in eyes with ciliary sulcus-implanted intraocular lenses (IOLs).
DESIGN: Retrospective cross-sectional study.
METHODS: One hundred thirty-two eyes of 132 consecutive patients who underwent phacoemulsification with implantation of a ZA9003 (Abbott Medical Optics Inc, Santa Ana, California, USA; 69 eyes) or AR40e (Abbott Medical Optics Inc; 63 eyes) IOL in the ciliary sulcus were enrolled. The modified effective lens position (ELP) equation based on the corneal radius (CR) (ELP - β0 - β1 × CR, where β0 and β1 are parameters in the linear regression analysis) was obtained using linear regression analysis in each IOL group, and was applied to the other IOL group in order to evaluate its effect on refractive outcomes in an independent data set. The median absolute error (MedAE) was predicted using the modified effective lens position equation and was compared with that predicted using the adjusted IOL power.
RESULTS: The modified effective lens position equation was ELP + 11.662 - 1.6225 × CR in the ZA9003 group and ELP + 10.606 - 1.4817 × CR in the AR40e group. The MedAE that was predicted using the modified effective lens position equation (0.34 diopter [D] in the ZA9003 group and 0.42 D in the AR40e group) was significantly smaller than that predicted using the adjusted IOL power (0.47 D in the ZA9003 group and 0.66 D in the AR40e group) (P < .001 and P = .005, respectively).
CONCLUSIONS: IOL power calculation using the modified Haigis formula effective lens position equation improved refractive outcomes in eyes with sulcus-implanted IOLs.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2015        PMID: 26456253     DOI: 10.1016/j.ajo.2015.09.040

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  4 in total

Review 1.  Evidence of Polyphenols Efficacy against Dry Eye Disease.

Authors:  Gaia Favero; Enrico Moretti; Kristína Krajčíková; Vladimíra Tomečková; Rita Rezzani
Journal:  Antioxidants (Basel)       Date:  2021-01-28

2.  Influence of corneal power on intraocular lens power of the second eye in the SRK/T formula in bilateral cataract surgery.

Authors:  Young Choi; Youngsub Eom; Jong Suk Song; Hyo Myung Kim
Journal:  BMC Ophthalmol       Date:  2017-12-28       Impact factor: 2.209

3.  AI-powered effective lens position prediction improves the accuracy of existing lens formulas.

Authors:  Tingyang Li; Joshua Stein; Nambi Nallasamy
Journal:  Br J Ophthalmol       Date:  2021-04-09       Impact factor: 5.908

4.  AI-Powered Effective Lens Position Prediction Improves the Accuracy of Existing Lens Formulas.

Authors:  Tingyang Li; Joshua D Stein; Nambi Nallasamy
Journal:  medRxiv       Date:  2020-11-03
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.