Evaluation of a bilinear model for attenuation correction using CT numbers generated from a parametric method.

L C Martinez1, A Calzado2.   

Abstract

A parametric model is used for the calculation of the CT number of some selected human tissues of known compositions (Hi) in two hybrid systems, one SPECT-CT and one PET-CT. Only one well characterized substance, not necessarily tissue-like, needs to be scanned with the protocol of interest. The linear attenuation coefficients of these tissues for some energies of interest (μ(i)) have been calculated from their tabulated compositions and the NIST databases. These coefficients have been compared with those calculated with the bilinear model from the CT number (μ(B)i). No relevant differences have been found for bones and lung. In the soft tissue region, the differences can be up to 5%. These discrepancies are attributed to the different chemical composition for the tissues assumed by both methods.
Copyright © 2015 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Attenuation correction; Hybrid systems; Stoichiometric calculation

Mesh:

Year:  2015        PMID: 26454019     DOI: 10.1016/j.apradiso.2015.09.014

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  2 in total

1.  Patient-specific image-based bone marrow dosimetry in Lu-177-[DOTA0,Tyr3]-Octreotate and Lu-177-DKFZ-PSMA-617 therapy: investigation of a new hybrid image approach.

Authors:  Astrid Gosewisch; Andreas Delker; Sebastian Tattenberg; Harun Ilhan; Andrei Todica; Julia Brosch; Lena Vomacka; Anika Brunegraf; Franz Josef Gildehaus; Sibylle Ziegler; Peter Bartenstein; Guido Böning
Journal:  EJNMMI Res       Date:  2018-08-03       Impact factor: 3.138

Review 2.  PET/MRI attenuation estimation in the lung: A review of past, present, and potential techniques.

Authors:  Joseph Lillington; Ludovica Brusaferri; Kerstin Kläser; Karin Shmueli; Radhouene Neji; Brian F Hutton; Francesco Fraioli; Simon Arridge; Manuel Jorge Cardoso; Sebastien Ourselin; Kris Thielemans; David Atkinson
Journal:  Med Phys       Date:  2020-01-01       Impact factor: 4.071

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.