| Literature DB >> 26451143 |
Claudio Casella1, Pierluigi Rossini2, Carlo Cappelli3, Chiara Nessi1, Riccardo Nascimbeni1, Nazario Portolani1.
Abstract
Background. A proper localisation of pathological parathyroid glands is essential for a minimally invasive approach in the surgical treatment of primary hyperparathyroidism (PHP). The recent introduction of portable mini gamma-cameras (pMGCs) enabled intraoperative scintigraphic scanning. The aim of our study is to evaluate the efficacy of this new method and compare it with the preoperative localisation surveys. Methods. 20 patients were studied; they were evaluated preoperatively by neck ultrasound and (99mm)Tc-sestaMIBI-scintigraphy and intraoperatively with the pMGC IP Guardian 2. The results obtained from the three evaluations were compared. Results. The pMGC presented a sensitivity of 95%, a specificity of 98.89%, and a diagnostic accuracy of 98.18%, which were higher than those of preoperative ultrasound (sensitivity 55%; specificity 95%; diagnostic accuracy 87%) and scintigraphy with (99mm)Tc-sestaMIBI (sensitivity 73.68%; specificity 96.05%; diagnostic accuracy 91.58%). Conclusions. The pMGC can be used effectively as an intraoperative method to find the correct location of the pathological parathyroid glands. The pMGC is more reliable than the currently used preoperative and intraoperative localisation techniques.Entities:
Year: 2015 PMID: 26451143 PMCID: PMC4586957 DOI: 10.1155/2015/134731
Source DB: PubMed Journal: Int J Endocrinol ISSN: 1687-8337 Impact factor: 3.257
Figure 1Intraoperative scans acquisition.
Figure 2Lower left adenoma (white arrow).
Patients with PHP.
| Age | |
| Average (IC 95%) | 63.7 (58.97–68.43) |
| Gender | |
| F | 13 (65%) |
| M | 7 (35%) |
| Serum calcium | |
| Average (IC 95%) | 12.26 mg/dL (11.49–13.03) |
| PTH | |
| Average (IC 95%) | 350.1 pg/mL (233.59–466.61) |
| Symptoms | |
| No | 6 (30%) |
| Yes | 14 (70%) |
| Kidney | 7 (35%) |
| Osteoarticular | 6 (30%) |
| Neuropsychic | 2 (10%) |
| Digestive | 1 (5%) |
| Acute hypercalcemia | 1 (5%) |
Results of preoperative and intraoperative surgical findings and pathological anatomical study locators.
| Ultrasound | Scintigraphy | Portable MGC | Surgical findings | Anatomy-pathological findings | |
|---|---|---|---|---|---|
| 1 | Left | Lower left | Lower left | Lower left | Lower left |
| 2 | Lower left | Lower left | Lower left | Lower left | Lower left |
| 3 | Lower left | Lower left | Lower left | Lower left | Lower left |
| 4 | Right | Middle right | Upper right | Upper right | Upper right |
| 5 | Middle left | Left | Upper left | Upper left | Upper left |
| 6 | No injuries | Lower left | Lower left | Lower left | Lower left |
| 7 | Lower right | Right | Lower right | Lower right | Lower right |
| 8 | Right | Paratracheal right | Paratracheal right | Paratracheal right | Paratracheal right |
| 9 | Middle left | Left | Middle left | Upper left | Upper left |
| 10 | Lower right | Lower right | Lower right | Lower right | Lower right |
| 11 | Lower right | Lower right | Lower right | Lower right | Lower right |
| 12 | Lower right | Lower right | Lower right | Lower right | Lower right |
| 13 | Middle right | Lower right | Lower right | Lower right | Lower right |
| 14 | Middle left | Lower left | Lower left | Lower left | Lower left |
| 15 | Upper left | Upper left | Upper left | Upper left | Upper left |
| 16 | Right | Upper left | Upper right | Upper right | Upper right |
| 17 | Lower left | Middle left | Lower left | Lower left | Lower left |
| 18 | Lower right | Lower right | Lower right | Lower right | Lower right |
| 19 | Upper left | Upper left | Upper left | Upper left | Upper left |
| 20 | Upper left | Middle left | Upper left | Upper left | Upper left |
Surgical findings.
| Lower left adenoma | 6 (30%) |
| Lower right adenoma | 6 (30%) |
| Upper left adenoma | 5 (25%) |
| Upper right adenoma | 2 (10%) |
| Ectopic adenoma | 1 (5%) |
Sensitivity, specificity, diagnostic accuracy, PPV, and NPV of the localisation investigations used according to the side of the neck.
| Sensitivity | Specificity | Diagnostic accuracy | PPV | NPV | |
|---|---|---|---|---|---|
| Ultrasound | 95% | 100% | 99% | 100% | 98.76% |
| Preoperative scintigraphy | 100% | 100% | 100% | 100% | 100% |
| pMGC | 100% | 100% | 100% | 100% | 100% |
Sensitivity, specificity, diagnostic accuracy, PPV, and NPV of the localisation used according to the quadrant of the neck.
| Sensitivity | Specificity | Diagnostic accuracy | PPV | NPV | |
|---|---|---|---|---|---|
| Ultrasound | 55% | 95% | 87% | 73.33% | 89.41% |
| Preoperative scintigraphy | 73.68% | 96.05% | 91.58% | 82.35% | 93.59% |
| pMGC | 95% | 98.89% | 98.18% | 95% | 98.88% |