Using motion parallax for laparoscopic surgery.

He Su1, Jianmin Li1, Huaifeng Zhang1, Jinhua Li1, Shuxin Wang1.   

Abstract

BACKGROUND: In typical stereo display systems, the reproduced 3D scene is distorted when the observer moves. Motion parallax is an important depth cue that has the ability to correct this distortion. More importantly, motion parallax enables the observer to look around objects to provide a better view.
METHODS: A robotically assisted laparoscope prototype was designed to provide motion parallax. A study to adjust the camera-head mapping ratio (gain of motion parallax) was performed. A series of phantom tests was conducted to test the effectiveness of motion parallax.
RESULTS: The experimental results showed that the motion parallax was effective, and the gain of motion parallax was subjective. For a regular laparoscope view distance, larger image zooming rates required smaller gain; for the same equivalent image size, further observer distance decreased the optimal gain.
CONCLUSIONS: Motion parallax could be used for improved visualization in laparoscopic surgery.
Copyright © 2015 John Wiley & Sons, Ltd. Copyright © 2015 John Wiley & Sons, Ltd.

Entities:  

Keywords:  3D display; laparoscopy; motion parallax; robotic surgery; stereo

Mesh:

Year:  2015        PMID: 26443699     DOI: 10.1002/rcs.1703

Source DB:  PubMed          Journal:  Int J Med Robot        ISSN: 1478-5951            Impact factor:   2.547


  1 in total

1.  Is 3D faster and safer than 4K laparoscopic cholecystectomy? A randomised-controlled trial.

Authors:  Matt Dunstan; Ralph Smith; Katie Schwab; Andrea Scala; Piers Gatenby; Martin Whyte; Tim Rockall; Iain Jourdan
Journal:  Surg Endosc       Date:  2019-07-18       Impact factor: 4.584