Comparison of positron emission tomography diffusion-weighted imaging (PET/DWI) registration quality in a PET/MR scanner: Zoomed DWI vs. Conventional DWI.

Koji Sagiyama1, Yuji Watanabe2, Ryotaro Kamei1, Shingo Baba1, Hiroshi Honda1.   

Abstract

PURPOSE: To compare zoomed diffusion-weighted imaging (z-DWI) with reduced field of view (FOV) by spatially selective radiofrequency pulses and conventional echo planar imaging (EPI) DWI (c-DWI) with regard to registration quality using positron emission tomography / magnetic resonance (PET/MR) in patients with malignant tumors.
MATERIALS AND METHODS: Fludeoxyglucose (18(F) ) PET imaging, c-DWI, and z-DWI were conducted simultaneously in 21 patients with known or suspected malignancy using a PET/MR system. A fusion image showing the largest tumor area was generated for analysis. Registration accuracy between PET and DWI was assessed based on the area of maximum overlap and central point displacement of the tumor. EPI factor, echo time (TE), matching area, and displacement were compared between c-DWI and z-DWI by paired t-test. Agreement of apparent diffusion coefficient (ADC) acquired by the two sequences were also assessed with linear regression s and Bland-Altman plot analysis.
RESULTS: Thirty-two lesions were detected on both PET and DWI (mean size 536.3 ± 471.8 mm(2) ). At least one lesion was found in all subjects. In all cases, EPI factor was smaller with z-DWI than c-DWI (43.1 ± 15.6 vs. 62.0 ± 10.0, P < 0.0001), and TE was also shorter for z-DWI (53.6 ± 3.6 msec vs. 65.2 ± 3.6 msec, P < 0.0001). Registration accuracy was better with z-DWI in 30 of 32 lesions (93.8%), and both average matching area and central point displacement were significantly improved (79.8 ± 18.1% vs. 61.8 ± 22.9%, P < 0.0001 and 3.92 ± 2.69 mm vs. 7.51 ± 4.07 mm, P < 0.0001). ADC values calculated with c-DWI and z-DWI showed good agreement.
CONCLUSION: Zoomed DWI reduces image distortion and provides better registration accuracy with PET images.
© 2015 Wiley Periodicals, Inc.

Entities:  

Keywords:  PET/MR; image fusion; spatially selective RF pulse; zoomed-EPI DWI

Mesh:

Substances:

Year:  2015        PMID: 26440585     DOI: 10.1002/jmri.25059

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  2 in total

1.  Multiparametric voxel-based analyses of standardized uptake values and apparent diffusion coefficients of soft-tissue tumours with a positron emission tomography/magnetic resonance system: Preliminary results.

Authors:  Koji Sagiyama; Yuji Watanabe; Ryotaro Kamei; Sungtak Hong; Satoshi Kawanami; Yoshihiro Matsumoto; Hiroshi Honda
Journal:  Eur Radiol       Date:  2017-06-20       Impact factor: 5.315

Review 2.  FDG Whole-Body PET/MRI in Oncology: a Systematic Review.

Authors:  Hyun Woo Kwon; Ann-Katharina Becker; Jin Mo Goo; Gi Jeong Cheon
Journal:  Nucl Med Mol Imaging       Date:  2016-04-07
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.