Temperature and time variations during osteotomies performed with different piezosurgical devices: an in vitro study.

R A Delgado-Ruiz1, D Sacks2, A Palermo3, J L Calvo-Guirado4, C Perez-Albacete4, G E Romanos5.   

Abstract

AIM: The aim of this experimental in vitro study was to evaluate the effects of the piezoelectric device in temperature and time variations in standardized osteotomies performed with similar tip inserts in bovine bone blocks.
METHODS: Two different piezosurgical devices were used the OE-F15(®) (Osada Inc., Los Angeles, California, USA) and the Surgybone(®) (Silfradent Inc., Sofia, Forli Cesena, Italy). Serrated inserts with similar geometry were coupled with each device (ST94 insert/test A and P0700 insert/test B). Osteotomies 10 mm long and 3 mm deep were performed in bone blocks resembling type II (dense) and type IV (soft) bone densities with and without irrigation. Thermal changes and time variations were recorded. The effects of bone density, irrigation, and device on temperature changes and time necessary to accomplish the osteotomies were analyzed.
RESULTS: Thermal analysis showed significant higher temperatures during piezosurgery osteotomies in hard bone without irrigation (P < 0.05). The type of piezosurgical device did not influence thermal variations (P > 0.05). Time analysis showed that the mean time values necessary to perform osteotomies were shorter in soft bone than in dense bone (P < 0.05).
CONCLUSIONS: Within the limitations of this in vitro study, it may be concluded that the temperature increases more in piezosurgery osteotomies in dense bone without irrigation; the time to perform the osteotomy with piezosurgery is shorter in soft bone compared to hard bone; and the piezosurgical device have a minimal influence in the temperature and time variations when a similar tip design is used during piezosurgery osteotomies.
© 2015 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  osteotomy; piezosurgery; temperature changes; time of surgery

Mesh:

Year:  2015        PMID: 26439590     DOI: 10.1111/clr.12709

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  4 in total

1.  Comparison of heat production and bone architecture changes in the implant site preparation with compressive osteotomes, osseodensification technique, piezoelectric devices, and standard drills: an ex vivo study on porcine ribs.

Authors:  Nishith Bhargava; Vittoria Perrotti; Vito Carlo Alberto Caponio; Victor Haruo Matsubara; Diana Patalwala; Alessandro Quaranta
Journal:  Odontology       Date:  2022-07-19       Impact factor: 2.885

2.  Temperature Values Variability in Piezoelectric Implant Site Preparation: Differences between Cortical and Corticocancellous Bovine Bone.

Authors:  Luca Lamazza; Girolamo Garreffa; Domenica Laurito; Marco Lollobrigida; Luigi Palmieri; Alberto De Biase
Journal:  Biomed Res Int       Date:  2016-03-24       Impact factor: 3.411

3.  Comparison of peri-implant bone loss between conventional drilling with irrigation versus low-speed drilling without irrigation.

Authors:  H Pellicer-Chover; D Peñarrocha-Oltra; A Aloy-Prosper; J-C Sanchis-Gonzalez; M-A Peñarrocha-Diago; M Peñarrocha-Diago
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2017-11-01

4.  Ridge augmentation-The new field of computerized guided surgery: A technical note for minimal-invasive bone splitting.

Authors:  Vasilios Alevizakos; Gergo Mitov; Marcus Schiller; Constantin von See
Journal:  Clin Case Rep       Date:  2021-03-20