Comparative Analysis of Urinary Total Proteins by Bicinchoninic Acid and Pyrogallol Red Molybdate Methods.

Padma Yalamati1, Aparna Varma Bhongir2, Madhulatha Karra3, Sashidhar Rao Beedu3.   

Abstract

BACKGROUND: The concentration of total proteins in urine is a good index of renal function, but its determination is found to be unreliable. The pyrogallol red molybdate (PRM) method for urine total proteins is being widely used in most of the hospitals because of its high sensitivity, better precision and its practicability. Bicinchoninic acid method (BCA) is also used for protein estimation and there have been no studies comparing this method with the PRM method in human urine samples. BCA method overestimates the urinary protein concentration in the presence of interfering substances. After removing the interfering substances present in the human urine samples the results of BCA method were compared with the PRM method. AIM: The purpose of the study is to identify whether the results of urine total proteins by BCA method are comparable to PRM method and can be used as an alternative to the PRM method. SETTING AND
DESIGN: This is a cross-sectional study done on fresh urine specimens from the hospital laboratory, covering a wide range of protein concentrations.
MATERIAL AND METHODS: Fresh urine specimens covering a wide range of protein concentrations (urine dipstick: nil, trace, 1+, 2+ and ≥ 3+) of 36 patients were analysed by both the methods. STATISTICAL ANALYSIS: Imprecision was determined by repeated analysis study and Inaccuracy was assessed by comparing the results of the patient's urine samples by both the methods using correlation plots, Bland and Altman, and Passing and Bablok regression analyses.
RESULTS: The coefficient of variation and mean (SD) for the BCA method were 4.6% and 799.1 (882.5) mg/L and for the PRM method were 5.1% and 802.1 (911.9) mg/L. The Pearson correlation coefficient, r was 0.93 (p < 0.0001). Method agreement studies showed no significant constant and proportional bias between both the methods.
CONCLUSION: In urine which is subjected to removal of interfering substances, the BCA results are comparable to PRM method.

Entities:  

Keywords:  Bland and Altman analysis; Correlation; Passing and Bablok regression analysis; Urinary protein assays

Year:  2015        PMID: 26435938      PMCID: PMC4576529          DOI: 10.7860/JCDR/2015/13543.6313

Source DB:  PubMed          Journal:  J Clin Diagn Res        ISSN: 0973-709X


  22 in total

1.  Improved biuret procedure for routine determination of urinary total proteins in clinical proteinuria.

Authors:  E W Rice
Journal:  Clin Chem       Date:  1975-03       Impact factor: 8.327

2.  Determination of total urinary protein, combining lowry sensitivity and biuret specificity.

Authors:  K Doetsch; R H Gadsden
Journal:  Clin Chem       Date:  1973-10       Impact factor: 8.327

3.  A biuret method for determination of protein in normal urine.

Authors:  J Savory; P H Pu; F W Sunderman
Journal:  Clin Chem       Date:  1968-12       Impact factor: 8.327

4.  Three turbidimetric methods for determining total protein compared.

Authors:  H H Nishi; R J Elin
Journal:  Clin Chem       Date:  1985-08       Impact factor: 8.327

5.  Six methods for urinary protein compared.

Authors:  L A McElderry; I F Tarbit; A J Cassells-Smith
Journal:  Clin Chem       Date:  1982-02       Impact factor: 8.327

6.  Evaluation of the Bradford method for protein determination in body fluids.

Authors:  E Schleicher; O H Wieland
Journal:  J Clin Chem Clin Biochem       Date:  1978-09

7.  Normal values for 24-h urinary protein excretion: total and low molecular weight proteins with a sex-related difference.

Authors:  H Kawakami; T Murakami; T Kajii
Journal:  Clin Nephrol       Date:  1990-05       Impact factor: 0.975

8.  Turbidimetric measurement of total urinary proteins: a revised method.

Authors:  S Shahangian; P I Brown; K O Ash
Journal:  Am J Clin Pathol       Date:  1984-05       Impact factor: 2.493

9.  Measurement of protein using bicinchoninic acid.

Authors:  P K Smith; R I Krohn; G T Hermanson; A K Mallia; F H Gartner; M D Provenzano; E K Fujimoto; N M Goeke; B J Olson; D C Klenk
Journal:  Anal Biochem       Date:  1985-10       Impact factor: 3.365

10.  Urine protein concentration estimation for biomarker discovery.

Authors:  Hiten D Mistry; Kate Bramham; Andrew J Weston; Malcolm A Ward; Andrew J Thompson; Lucy C Chappell
Journal:  Pregnancy Hypertens       Date:  2013-08-05       Impact factor: 2.899

View more
  7 in total

1.  Usefulness of the quantitative measurement of urine protein at a community-based health checkup: a cross-sectional study.

Authors:  Masahiro Naruse; Masashi Mukoyama; Jun Morinaga; Masanobu Miyazaki; Kunitoshi Iseki; Kunihiro Yamagata
Journal:  Clin Exp Nephrol       Date:  2019-09-20       Impact factor: 2.801

2.  Overexpression of mRNA-decapping enzyme 1a affects survival rate in colorectal carcinoma.

Authors:  Chuanqing Wu; Weizhen Liu; Tuo Ruan; Xiaojie Zhu; Kaixiong Tao; Weikang Zhang
Journal:  Oncol Lett       Date:  2018-05-16       Impact factor: 2.967

3.  Indirubin-3'-monoxime suppresses amyloid-beta-induced apoptosis by inhibiting tau hyperphosphorylation.

Authors:  Shu-Gang Zhang; Xiao-Shan Wang; Ying-Dong Zhang; Qing Di; Jing-Ping Shi; Min Qian; Li-Gang Xu; Xing-Jian Lin; Jie Lu
Journal:  Neural Regen Res       Date:  2016-06       Impact factor: 5.135

4.  Synergic Effect of α-Mangostin on the Cytotoxicity of Cisplatin in a Cervical Cancer Model.

Authors:  Jazmin M Pérez-Rojas; Raquel González-Macías; Jaime González-Cortes; Rafael Jurado; José Pedraza-Chaverri; Patricia García-López
Journal:  Oxid Med Cell Longev       Date:  2016-12-08       Impact factor: 6.543

5.  Lower Serum Vitamin D Metabolite Levels in Relation to Circulating Cytokines/Chemokines and Metabolic Hormones in Pregnant Women with Hypertensive Disorders.

Authors:  Ramu Adela; Roshan M Borkar; Navneeta Mishra; Murali Mohan Bhandi; Gayatri Vishwakarma; B Aparna Varma; Srinivas Ragampeta; Sanjay K Banerjee
Journal:  Front Immunol       Date:  2017-03-13       Impact factor: 7.561

6.  Erythropoietin Mimetic Peptide (pHBSP) Corrects Endothelial Dysfunction in a Rat Model of Preeclampsia.

Authors:  Mikhail Korokin; Vladimir Gureev; Oleg Gudyrev; Ivan Golubev; Liliya Korokina; Anna Peresypkina; Tatiana Pokrovskaia; Galina Lazareva; Vladislav Soldatov; Mariya Zatolokina; Anna Pobeda; Elena Avdeeva; Evgeniya Beskhmelnitsyna; Tatyana Denisyuk; Natalia Avdeeva; Olga Bushueva; Mikhail Pokrovskii
Journal:  Int J Mol Sci       Date:  2020-09-15       Impact factor: 5.923

7.  Effect of bovine lactoferrin as a novel therapeutic agent in a rat model of sepsis-induced acute lung injury.

Authors:  Nannan Han; Hengjie Li; Gang Li; Ye Shen; Min Fei; Yong Nan
Journal:  AMB Express       Date:  2019-10-31       Impact factor: 3.298

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.