Modification of a commercial gas chromatography isotope ratio mass spectrometer for on-line carbon isotope dilution: Evaluation of its analytical characteristics for the quantification of organic compounds.

Laura Alonso Sobrado1, Mario Robledo Fernández1, Sergio Cueto Díaz1, Jorge Ruiz Encinar2, J Ignacio García Alonso3.   

Abstract

We describe the instrumental modification of a commercial gas chromatography isotope ratio mass spectrometer (GC-IRMS) and its application for on-line carbon isotope dilution. The main modification consisted in the addition of a constant flow of enriched (13)CO2 diluted in helium after the chromatographic column through the splitter holder located inside the chromatographic oven of the instrument. In addition, and in contrast to the conventional mode of operation of GC-IRMS instruments where the signal at m/z 45 is amplified 100-fold with respect to the signal at m/z 44, the same signal amplification was used in both Faraday cups at m/z 44 and 45. Under these conditions isotope ratio precision for the ratio 44/45 was around 0.05% RSD (n=50). The evaluation of the instrument was performed with mixtures of organic compounds including 11 n-alkanes, 16 PAHs, 12 PCBs and 3 benzothiophenes. It was observed that compounds of very different boiling points could be analysed without discrimination in the injector when a Programmable Temperature Vaporizer (PTV) injector was employed. Moreover, the presence of heteroatoms (Cl or S) in the structure of the organic compounds did not affect their combustion efficiency and therefore the trueness of the results. Quantitative results obtained for all the analytes assayed were excellent in terms of precision (<3% RSD) and accuracy (average relative error≤4%) and what is more important using a single and simple generic internal standard for quantification.
Copyright © 2015 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Combustion; Compound-independent calibration; Gas chromatography; Mass spectrometry; On-line isotope dilution

Mesh:

Substances:

Year:  2015        PMID: 26435309     DOI: 10.1016/j.chroma.2015.09.071

Source DB:  PubMed          Journal:  J Chromatogr A        ISSN: 0021-9673            Impact factor:   4.759


  1 in total

1.  Systematic comparison of post-column isotope dilution using LC-CO-IRMS with qNMR for amino acid purity determination.

Authors:  Philip J H Dunn; Dmitry Malinovsky; Eli Achtar; Cailean Clarkson; Heidi Goenaga-Infante
Journal:  Anal Bioanal Chem       Date:  2019-09-12       Impact factor: 4.142

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.