Literature DB >> 26427913

The impact of venous occlusion per se on forearm muscle blood flow: implications for the near-infrared spectroscopy venous occlusion technique.

Troy J Cross1,2, Surendran Sabapathy1.   

Abstract

The purpose of this study was to examine the effect of venous occlusion per se on forearm muscle blood flow, as determined by the near-infrared spectroscopy (NIRS) venous occlusion technique (NIRS-VOT). NIRS data were obtained from the flexor digitorum superficialis (FDS) muscle on the dominant arm of 16 young, ostensibly healthy participants (14 men and two women; 30 ± 6 year; 73 ± 7 kg). Participants completed a series of five venous occlusion trials while seated at rest, and a series of 12 venous occlusion trials during a reactive hyperaemia induced by 5 min of forearm arterial occlusion. The NIRS-VOT was used to assess FDS muscle blood flow (Q˙mus), beat-by-beat, over the first four cardiac beats during venous occlusions. Q˙mus was also reported as a cumulative value, wherein the first two, first three and first four cardiac beats were used to calculate muscle blood flow. We observed that Q˙mus was highest when calculated over the first cardiac beat during venous occlusions performed at rest and throughout reactive hyperaemia (P<0·05). Moreover, the inclusion of more than one cardiac beat in the calculation of Q˙mus underestimated muscle blood flows, irrespective of the prevailing level of arterial inflow. These findings support the idea that venous occlusion per se affects the measurement of Q˙mus via the NIRS-VOT. Accordingly, it is recommended that Q˙mus is determined over the first cardiac beat when using the NIRS-VOT to assess microvascular blood flow of human forearm muscles.
© 2015 Scandinavian Society of Clinical Physiology and Nuclear Medicine. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  forearm; muscle blood flow; near-infrared spectroscopy; venous occlusion

Mesh:

Year:  2015        PMID: 26427913     DOI: 10.1111/cpf.12301

Source DB:  PubMed          Journal:  Clin Physiol Funct Imaging        ISSN: 1475-0961            Impact factor:   2.273


  6 in total

1.  Cuff inflation time significantly affects blood flow recorded with venous occlusion plethysmography.

Authors:  Rehan T Junejo; Clare J Ray; Janice M Marshall
Journal:  Eur J Appl Physiol       Date:  2019-01-08       Impact factor: 3.078

2.  Measurement of muscle blood flow and O2 uptake via near-infrared spectroscopy using a novel occlusion protocol.

Authors:  Joshua J Dennis; Chad C Wiggins; Joshua R Smith; Jennifer M J Isautier; Bruce D Johnson; Michael J Joyner; Troy J Cross
Journal:  Sci Rep       Date:  2021-01-13       Impact factor: 4.379

3.  Time-domain NIRS system based on supercontinuum light source and multi-wavelength detection: validation for tissue oxygenation studies.

Authors:  Aleh Sudakou; Frédéric Lange; Helene Isler; Pranav Lanka; Stanislaw Wojtkiewicz; Piotr Sawosz; Daniel Ostojic; Martin Wolf; Antonio Pifferi; Ilias Tachtsidis; Adam Liebert; Anna Gerega
Journal:  Biomed Opt Express       Date:  2021-09-30       Impact factor: 3.732

4.  Evidence that large vessels do affect near infrared spectroscopy.

Authors:  Stefano Seddone; Leonardo Ermini; Piero Policastro; Luca Mesin; Silvestro Roatta
Journal:  Sci Rep       Date:  2022-02-09       Impact factor: 4.379

5.  Comparison of systemic and peripheral responses during high-intensity interval exercise under voluntary hypoventilation vs. hypoxic conditions.

Authors:  Ayano Imai; Keiichi Yamaguchi; Kazushige Goto
Journal:  Phys Act Nutr       Date:  2022-06-30

6.  The Effects of Uniquely-Processed Titanium on Balance and Walking Performance in Healthy Older Adults.

Authors:  Melissa J Black; Adam A Lucero; Philip W Fink; Lee Stoner; Sarah P Shultz; Sally D Lark; David S Rowlands
Journal:  J Funct Biomater       Date:  2018-06-08
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.