CLINICAL AUDIT OF IMAGE QUALITY IN RADIOLOGY USING VISUAL GRADING CHARACTERISTICS ANALYSIS.

Erik Tesselaar1, Nils Dahlström2, Michael Sandborg3.   

Abstract

The aim of this work was to assess whether an audit of clinical image quality could be efficiently implemented within a limited time frame using visual grading characteristics (VGC) analysis. Lumbar spine radiography, bedside chest radiography and abdominal CT were selected. For each examination, images were acquired or reconstructed in two ways. Twenty images per examination were assessed by 40 radiology residents using visual grading of image criteria. The results were analysed using VGC. Inter-observer reliability was assessed. The results of the visual grading analysis were consistent with expected outcomes. The inter-observer reliability was moderate to good and correlated with perceived image quality (r(2) = 0.47). The median observation time per image or image series was within 2 min. These results suggest that the use of visual grading of image criteria to assess the quality of radiographs provides a rapid method for performing an image quality audit in a clinical environment.
© The Author 2015. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

Mesh:

Year:  2015        PMID: 26410763     DOI: 10.1093/rpd/ncv411

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  3 in total

1.  Visual grading analysis of digital neonatal chest phantom X-ray images: Impact of detector type, dose and image processing on image quality.

Authors:  M H Smet; L Breysem; E Mussen; H Bosmans; N W Marshall; L Cockmartin
Journal:  Eur Radiol       Date:  2018-02-19       Impact factor: 5.315

2.  European guidelines on quality criteria for diagnostic radiographic images of the lumbar spine - an intra- and inter-observer reproducibility study.

Authors:  Klaus Doktor; Maria Lind Vilholm; Aldis Hardardóttir; Henrik Wulff Christensen; Jens Lauritsen
Journal:  Chiropr Man Therap       Date:  2019-05-01

3.  Comparison of T1wFLAIR and T1wTSE sequences in imaging the brain of small animals using high-field MRI.

Authors:  Chiara Bergamino; Séamus Hoey; Kenneth Waller; Cliona Skelly
Journal:  Ir Vet J       Date:  2019-07-06       Impact factor: 2.146