Dedicated mobile volumetric cone-beam computed tomography for human brain imaging: A phantom study.

Jong-Hyun Ryu1, Tae-Hoon Kim1, Chang-Won Jeong1, Hong-Young Jun1, Dong-Woon Heo1, Jinseok Lee1, Kyong-Woo Kim2, Kwon-Ha Yoon1,3.   

Abstract

BACKGROUND: Mobile computed tomography (CT) with a cone-beam source is increasingly used in the clinical field. Mobile cone-beam CT (CBCT) has great merits; however, its clinical utility for brain imaging has been limited due to problems including scan time and image quality.
OBJECTIVE: The aim of this study was to develop a dedicated mobile volumetric CBCT for obtaining brain images, and to optimize the imaging protocol using a brain phantom.
METHODS: The mobile volumetric CBCT system was evaluated with regards to scan time and image quality, measured as signal-to-noise-ratio (SNR), contrast-to-noise-ratio (CNR), spatial resolution (10% MTF), and effective dose. Brain images were obtained using a CT phantom.
RESULTS: The CT scan took 5.14 s at 360 projection views. SNR and CNR were 5.67 and 14.5 at 120 kV/10 mA. SNR and CNR values showed slight improvement as the x-ray voltage and current increased (p < 0.001). Effective dose and 10% MTF were 0.92 mSv and 360 μ m at 120 kV/10 mA. Various intracranial structures were clearly visible in the brain phantom images.
CONCLUSIONS: Using this CBCT under optimal imaging acquisition conditions, it is possible to obtain human brain images with low radiation dose, reproducible image quality, and fast scan time.

Entities:  

Keywords:  Cone-beam computed tomography; brain imaging; dedicated mobile CT; volumetric computed tomography

Mesh:

Year:  2015        PMID: 26410658     DOI: 10.3233/XST-150502

Source DB:  PubMed          Journal:  J Xray Sci Technol        ISSN: 0895-3996            Impact factor:   1.535


  2 in total

1.  Development of Portable Digital Radiography System with a Device for Monitoring X-ray Source-Detector Angle and Its Application in Chest Imaging.

Authors:  Tae-Hoon Kim; Dong-Woon Heo; Chang-Won Jeong; Jong-Hyun Ryu; Hong Young Jun; Seung-Jun Han; Taeuk Ha; Kwon-Ha Yoon
Journal:  Sensors (Basel)       Date:  2017-03-07       Impact factor: 3.576

2.  Mobile computed tomography for the efficient allocation of medical resources in patients with COVID-19 pneumonia: A case report.

Authors:  Jae Wan Jung; Hak Ryul Kim; Kwon Ha Yoon; Chul Park
Journal:  Medicine (Baltimore)       Date:  2021-11-19       Impact factor: 1.889

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.