Performance evaluation of the Sysmex XN-1000 hematology analyzer in assessment of the white blood cell count differential in pediatric specimens.

P-H Becker1, O Fenneteau1, L Da Costa1,2,3,4.   

Abstract

INTRODUCTION: The automated XN-1000 hematology analyzer enables to perform a blood cell count and a leukocyte differential. When abnormal cells were detected, a flag was generated by the analyzer and a manual microscopic examination of the corresponding blood film was performed.
METHODS: We compared the white blood cell differentials provided by the automated hematology analyzer XN-1000 in a pediatric population (n = 765) with those obtained through microscopic examination by cytologists and those obtained using a previous version of this analyzer, the XE-2100. Leukocytes count as well as flags sensitivity and specificity was analyzed.
RESULTS: The leukocytes count provided by the analyzer is in good accordance with the differential obtained by manual count in children older than 3 months. The sensitivity for blast detection is 99% and the detection of reactive cells is 63%. The flag specificity remains low (<35%) for blood samples collected from infants between 8 days and 2 years of age, but increases up to 67% thereafter. The results obtained with the XN-1000 analyzer show an improvement in comparison with those obtained with the XE-2100 analyzer.
CONCLUSION: The automated WBC differential provided by the XN-1000 analyzer in the pediatric setting is accurate, but a meticulous microscopic examination of blood smears remains necessary for infants up to 3 months of age to validate the analyzer flags.
© 2015 John Wiley & Sons Ltd.

Entities:  

Keywords:  Automated blood cells analyzers; Sysmex XN-1000; WBC differentials; flags; pediatrics

Mesh:

Year:  2015        PMID: 26407903     DOI: 10.1111/ijlh.12436

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  7 in total

1.  Comparing automated vs manual leukocyte differential counts for quantifying the 'left shift' in the blood of neonates.

Authors:  B C MacQueen; R D Christensen; B A Yoder; E Henry; V L Baer; S T Bennett; H M Yaish
Journal:  J Perinatol       Date:  2016-06-09       Impact factor: 2.521

2.  Caesarean delivery, immune function and inflammation in early life among Ecuadorian infants and young children.

Authors:  A L Thompson
Journal:  J Dev Orig Health Dis       Date:  2019-02-07       Impact factor: 2.401

3.  Overall equipment effectiveness, efficiency and slide review analysis of high-end hematology analyzers.

Authors:  Shubham Rastogi; Sukesh Chandran Nair; Pandiyan Murugan; Asady Sukanya Sukumar; Joy J Mammen; Saravanan Mullai
Journal:  Pract Lab Med       Date:  2022-04-18

4.  A Multidimensional Electronic Hydroxyurea Adherence Intervention for Children With Sickle Cell Disease: Single-Arm Before-After Study.

Authors:  Susan Creary; Deena Chisolm; Joseph Stanek; Jane Hankins; Sarah H O'Brien
Journal:  JMIR Mhealth Uhealth       Date:  2019-08-08       Impact factor: 4.773

5.  The Sysmex XN-L (XN-350) hematology analyzer offers a compact solution for laboratories in niche diagnostics.

Authors:  Tania A Khartabil; Magda M de Frankrijker; Yolanda B de Rijke; Henk Russcher
Journal:  Int J Lab Hematol       Date:  2020-09-19       Impact factor: 2.877

6.  Evaluation of automated hematology analyzer DYMIND DH76 compared to SYSMEX XN 1000 system.

Authors:  Milena Velizarova; Teodora Yacheva; Mariana Genova; Dobrin Svinarov
Journal:  J Med Biochem       Date:  2021-09-03       Impact factor: 3.402

7.  Performance evaluation of the XPEN60 CRP&SAA, a novel automated hematology analyzer, in detecting complete blood count, C-reactive protein, and serum amyloid A.

Authors:  Zengyan Zong; Lijuan Kan; Yanping Luo; Cheng Wang; Xibu Wang; Mengmeng Wang; Dan Xiong; Xiuming Zhang
Journal:  J Clin Lab Anal       Date:  2020-06-09       Impact factor: 2.352

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.