Scattered radiation doses absorbed by technicians at different distances from X-ray exposure: Experiments on prosthesis.

Hsien-Wen Chiang1, Ya-Ling Liu2,3, Tou-Rong Chen4, Chun-Lon Chen1, Hsien-Jen Chiang1, Shin-Yu Chao1.   

Abstract

UNLABELLED: This work aimed to investigate the spatial distribution of scattered radiation doses induced by exposure to the portable X-ray, the C-arm machine, and to simulate the radiologist without a shield of lead clothing, radiation doses absorbed by medical staff at 2 m from the central exposure point. MATERIAL AND
METHOD: With the adoption of the Rando Phantom, several frequently X-rayed body parts were exposed to X-ray radiation, and the scattered radiation doses were measured by ionization chamber dosimeters at various angles from the patient. Assuming that the central point of the X-ray was located at the belly button, five detection points were distributed in the operation room at 1 m above the ground and 1-2 m from the central point horizontally.
RESULTS: The radiation dose measured at point B was the lowest, and the scattered radiation dose absorbed by the prosthesis from the X-ray's vertical projection was 0.07 ±0.03 μGy, which was less than the background radiation levels. The Fluke biomedical model 660-5DE (400 cc) and 660-3DE (4 cc) ion chambers were used to detect air dose at a distance of approximately two meters from the central point. The AP projection radiation doses at point B was the lowest (0.07±0.03 μGy) and the radiation doses at point D was the highest (0.26±0.08 μGy) .Only taking the vertical projection into account, the radiation doses at point B was the lowest (0.52 μGy), and the radiation doses at point E was the highest (4 μGy).The PA projection radiation at point B was the lowest (0.36 μGy) and the radiation doses at point E was the highest(2.77 μGy), occupying 10-32% of the maximum doses. The maximum dose in five directions was nine times to the minimum dose. When the PX and the C-arm machine were used, the radiation doses at a distance of 2 m were attenuated to the background radiation level. The radiologist without a lead shield should stand at point B of patient's feet. Accordingly, teaching materials on radiation safety for radiological interns and clinical technicians were formulated.

Entities:  

Keywords:  Portable X-ray machine; dose; exposure; ionization chamber dosimeter

Mesh:

Year:  2015        PMID: 26405929     DOI: 10.3233/BME-151463

Source DB:  PubMed          Journal:  Biomed Mater Eng        ISSN: 0959-2989            Impact factor:   1.300


  1 in total

1.  Evaluation of scattered radiation dose received by medical staff during uterine artery embolization in the operating room.

Authors:  Hsien-Wen Chiang; Hsien-Jen Chiang; Jung-Hui Li; Leo Leung-Chit Tsang
Journal:  Technol Health Care       Date:  2020       Impact factor: 1.285

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.