Literature DB >> 26386167

Effect of processing, sterilization and crosslinking on UHMWPE fatigue fracture and fatigue wear mechanisms in joint arthroplasty.

Farzana Ansari1, Michael D Ries2, Lisa Pruitt3.   

Abstract

Ultra high molecular weight polyethylene (UHMWPE) has been used as a bearing surface in total joint replacements (TJR) for nearly five decades. This semi-crystalline polymer has extraordinary energetic toughness owing to its high molecular weight and entanglement density. However, it is challenged by a need to offer a combined resistance to fatigue, wear and oxidation in vivo. The processing, sterilization treatment, and microstructural tailoring of UHMWPE has evolved considerably in the past 50 years but an optimized microstructure remains elusive. This review seeks to provide an overview of this processing history to address two primary questions: First, how does microstructure affect fatigue fracture and fatigue wear mechanisms in UHMWPE? And second, can microstructure be optimized to provide resistance to fatigue, oxidation and wear in vivo? Previous literature demonstrates that while crosslinking improves resistance to adhesive/abrasive wear, it also reduces resistance to fatigue crack propagation and fatigue wear by restricting molecular mobility and rendering the polymer more brittle. Crystallinity improves fatigue resistance but generally increases elastic modulus and concomitant contact stresses in vivo. The presence of fusion defects or oxidation reduces further fatigue resistance and enhances fatigue wear. Thus, UHMWPE microstructural evolution comes with trade-offs. Currently there is no singular formulation of UHMWPE that is ideal for all TJR applications.
Copyright © 2015 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Crosslinking; Oxidation; Total joint replacement; UHMWPE fracture; Ultrahigh molecular weight polyethylene

Mesh:

Substances:

Year:  2015        PMID: 26386167     DOI: 10.1016/j.jmbbm.2015.08.026

Source DB:  PubMed          Journal:  J Mech Behav Biomed Mater        ISSN: 1878-0180


  6 in total

1.  Equivalent wear performance of dual mobility bearing compared with standard bearing in total hip arthroplasty: in vitro study.

Authors:  Gaël Gaudin; André Ferreira; Romain Gaillard; Jean Louis Prudhon; Jacques H Caton; Sébastien Lustig
Journal:  Int Orthop       Date:  2016-11-22       Impact factor: 3.075

2.  Mid-term clinical outcomes and survivorship of medial-pivot total knee arthroplasty-a mean five year follow-up based on one thousand, one hundred and twenty eight cases.

Authors:  Shuai Xiang; Yingzhen Wang; Chengyu Lv; Changyao Wang; Haining Zhang
Journal:  Int Orthop       Date:  2021-03-30       Impact factor: 3.075

3.  Highly Crosslinked Polyethylene Liners Have Negligible Wear at 10 Years: A Radiostereometric Analysis Study.

Authors:  David G Campbell; Stuart A Callary
Journal:  Clin Orthop Relat Res       Date:  2022-03-01       Impact factor: 4.755

4.  Evaluation of double-high insert mid-term outcomes in cruciate-retaining medial-pivotal total knee arthroplasty - a propensity score-matched analysis with averaged 8-year follow-up.

Authors:  Wenzhe Wang; Shuai Xiang; Yingzhen Wang; Chengyu Lv; Changyao Wang; Haining Zhang
Journal:  BMC Musculoskelet Disord       Date:  2022-06-14       Impact factor: 2.562

5.  Does the Hirsch Index Improve Research Quality in the Field of Biomaterials? A New Perspective in the Biomedical Research Field.

Authors:  Saverio Affatato; Massimiliano Merola
Journal:  Materials (Basel)       Date:  2018-10-13       Impact factor: 3.623

6.  Chemerin/ChemR23 signaling mediates the effects of ultra-high molecular weight polyethylene wear particles on the balance between osteoblast and osteoclast differentiation.

Authors:  Fengchao Zhao; Dingwei Cang; Jianzhi Zhang; Li Zheng
Journal:  Ann Transl Med       Date:  2021-07
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.