Literature DB >> 26385495

Analytical and clinical validation of parathyroid hormone (PTH) measurement in fine-needle aspiration biopsy (FNAB) washings.

Hemamalini Ketha1, Michael A Lasho1, Alicia Algeciras-Schimnich2.   

Abstract

BACKGROUND: Parathyroid hormone (PTH) quantitation in fine needle aspirate biopsy (FNAB) saline washings complements current modalities for parathyroid tissue localization.
OBJECTIVES: To establish the performance characteristics of the Roche Elecsys intact PTH immunoassay in FNAB needle washings and its diagnostic performance for the identification of parathyroid tissue. DESIGN AND METHODS: Accuracy, precision, reportable range, and analytical specificity and sensitivity for the intact PTH immunoassay in FNAB needle washings were established. For clinical validation, 93 specimens from 79 patients were evaluated. Diagnostic cut-offs were established via receiver operator characteristic (ROC) curve analysis. Performance of PTH in FNAB needle washings was compared to cytology.
RESULTS: Measurement of the PTH in FNAB needle washings demonstrated a matrix interference that was overcome by supplementation of the samples with a protein based matrix prior to analysis. ROC area under the curve (AUC) was 0.96 for PTH in FNAB needle washings. A PTH concentration ≥100pg/mL showed 100% specificity and 82% sensitivity for identifying parathyroid tissue. On histology-confirmed parathyroid specimens, 21/38 (55%) were correctly identified by cytology; whereas 31/38 (82%) were identified by PTH.
CONCLUSIONS: Measurement of PTH in FNAB washings complements cytology for identification of parathyroid tissue. Analytical validation to exclude interference in the PTH immunoassay and proper localization of the parathyroid tissue by ultrasound is necessary to ensure the robustness of the method.
Copyright © 2015 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  FNAB needle washings; Fine needle aspirate biopsy (FNAB); Parathyroid hormone (PTH)

Mesh:

Substances:

Year:  2015        PMID: 26385495     DOI: 10.1016/j.clinbiochem.2015.09.006

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  2 in total

1.  Evaluation of Autofluorescence in Identifying Parathyroid Glands by Measuring Parathyroid Hormone in Fine-Needle Biopsy Washings.

Authors:  Zhen Liu; Run-Sheng Ma; Jun-Li Jia; Tao Wang; Dao-Hong Zuo; De-Tao Yin
Journal:  Front Endocrinol (Lausanne)       Date:  2022-01-21       Impact factor: 5.555

2.  The Current Role of Parathyroid Fine-Needle Biopsy (P-FNAB) with iPTH-Washout Concentration (iPTH-WC) in Primary Hyperparathyroidism: A Single Center Experience and Literature Review.

Authors:  Łukasz Obołończyk; Izabela Karwacka; Piotr Wiśniewski; Krzysztof Sworczak; Tomasz Osęka
Journal:  Biomedicines       Date:  2022-01-06
  2 in total