Incidental bronchial injury by soft coagulation.

Tomoki Shibano1, Shunsuke Endo1, Shinichi Otani1, Tomoyuki Nakano1.   

Abstract

Soft coagulation is a hemostat system of electrosurgical units, which automatically regulates its output voltage below 200 V, to avoid excessive output that causes carbonization of the target tissue. However, this new minimally invasive technology still has the potential risk of tissue damage during surgery. We encountered three patients with bronchial injury caused by the above system; one of whom had bronchopleural fistula. This is believed to be the first report emphasizing the adverse effects of the soft coagulation system in thoracic surgery, giving a warning to the application of this convenient device.

Entities:  

Keywords:  Thoracic surgery; bronchial fistula; equipment

Year:  2015        PMID: 26380775      PMCID: PMC4561262          DOI: 10.3978/j.issn.2072-1439.2015.08.05

Source DB:  PubMed          Journal:  J Thorac Dis        ISSN: 2072-1439            Impact factor:   2.895


  4 in total

1.  Efficacy of SOFT COAG for intraoperative bleeding in thoracic surgery.

Authors:  Tohru Sakuragi; Hiroko Ohma; Hitoshi Ohteki
Journal:  Interact Cardiovasc Thorac Surg       Date:  2009-08-27

2.  Hypertensive crisis during hemostasis of adrenal bleeding using a soft-coagulation device.

Authors:  Taketo Kawai; Motofumi Suzuki; Masafumi Otsuka; Shusaku Nakazono; Yoshinori Tanaka; Yukio Homma
Journal:  Int J Urol       Date:  2015-04-14       Impact factor: 3.369

3.  VIO soft-coagulation system for major pulmonary resections: results in 68 patients with primary lung cancer.

Authors:  Akihiko Uchiyama; Kei Miyoshi; Katsuya Nakamura
Journal:  Gen Thorac Cardiovasc Surg       Date:  2011-03-30

4.  Morphological changes induced by extensive endobronchial electrocautery.

Authors:  C Verkindre; A Brichet; C A Maurage; P Ramon; J P Homasson; C H Marquette
Journal:  Eur Respir J       Date:  1999-10       Impact factor: 16.671

  4 in total
  3 in total

1.  Usefulness of a suction ball coagulation probe for hemostasis in complete VATS lobectomy for patients with non-small cell lung cancer.

Authors:  Hiroyuki Deguchi; Makoto Tomoyasu; Wataru Shigeeda; Yuka Kaneko; Hironaga Kanno; Eri Maeda; Hajime Saito
Journal:  Surg Today       Date:  2019-01-25       Impact factor: 2.549

2.  A new method using a vessel-sealing system provides coagulation effects to various types of bleeding with less thermal damage.

Authors:  Shosaburo Oyama; Takashi Nonaka; Keitaro Matsumoto; Daisuke Taniguchi; Yasumasa Hashimoto; Tomohiro Obata; Makoto Hisanaga; Masaaki Moriyama; Naoto Matsuo; Hideo Wada; Kiyoaki Hamada; Kouki Wakata; Tetsuro Tominaga; Shigekazu Hidaka; Terumitsu Sawai; Takeshi Nagayasu
Journal:  Surg Endosc       Date:  2020-10-15       Impact factor: 4.584

Review 3.  Advances and safe use of energy devices in lung cancer surgery.

Authors:  Takahiro Homma
Journal:  Gen Thorac Cardiovasc Surg       Date:  2022-02-02
  3 in total