PURPOSE: The purpose of our study was to investigate the potential pitfalls associated with different vessel segmentation methods using the built-in software of the Retinal Function Imager (RFI) for the analysis of retinal blood flow velocity (BFV).
METHODS: Ten eyes of nine healthy subjects were enrolled in the study. Retinal blood flow measurements were obtained with the RFI device with a 20° field of view imaging. The same grader segmented the retinal vasculature using the RFI software in both sessions, with segments ranging in length from 50 to 100 pixels ("short segments") and 100-200 pixels ("long segments"). The blood flow velocities for the arteriolar and venular system were calculated, and the percentage of excluded vessel segments with high coefficients of variation (>45 %) was recorded and compared by paired t test. Spearman's correlation was used to analyze the relationship between measurements by the two vessel segmentation methods.
RESULTS: The number of analyzed vessel segments did not differ significantly between the two groups (28.6 ± 2.6 short and 26.7 ± 4.6 long segments), while the percentage of acceptable segments was significantly higher in the long segment group (65.2 ± 11.4 % vs 85.2 ± 5.87 %, p = 0.001). In the short segment group, more than 15 % of vessel segments were rejected in all subjects, while in the long segment group only three subjects had a rejection rate of greater than 15 % (16.7 %, 18.7 % and 28 %). Both arteriolar and venular velocities were lower in the short segment group, although it reached significance only for arteriolar velocities (3.93 ± 0.55 vs. 4.45 ± 0.76 mm/s, p = 0.036 and 2.95 ± 0.56 vs. 3.17 ± 0.84 mm/s, p = 0.201 for arterioles and venules, respectively). Only venular velocities showed significant correlation (p = 0.003, R (2) = 0.67) between the two groups.
CONCLUSIONS: Our results suggest that BFV measurements by the RFI may be affected by segment length, and care should therefore be taken in choosing vessel segment lengths used during the analysis of RFI data. Long segments of 100-200 pixels (400-800 μm) seem to provide more robust measurements, which can be explained by the analysis methodology of the RFI device.

Entities:  

Keywords:  Image analysis; Reproducibility; Retinal Function Imager; Retinal blood flow

Mesh:

Year:  2015        PMID: 26373549      PMCID: PMC4792806          DOI: 10.1007/s00417-015-3166-0

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  26 in total

1.  Relationship between blood flow velocities in retrobulbar vessels and laser Doppler flowmetry at the optic disk in glaucoma patients.

Authors:  Z Bohdanecka; S Orgül; A B Meyer; C Prünte; J Flammer
Journal:  Ophthalmologica       Date:  1999       Impact factor: 3.250

2.  Initial report of quantification of retinal blood flow velocity in normal human subjects using the Retinal Functional Imager (RFI).

Authors:  Gennady Landa; Anisha A Jangi; Patricia M T Garcia; Richard B Rosen
Journal:  Int Ophthalmol       Date:  2012-04-08       Impact factor: 2.031

Review 3.  Special report: Noninvasive multi-parameter functional optical imaging of the eye.

Authors:  Darin A Nelson; Sara Krupsky; Ayala Pollack; Eyal Aloni; Michael Belkin; Ivo Vanzetta; Mordechai Rosner; Amiram Grinvald
Journal:  Ophthalmic Surg Lasers Imaging       Date:  2005 Jan-Feb

4.  New patterns of retinal collateral circulation are exposed by a retinal functional imager (RFI).

Authors:  G Landa; R B Rosen
Journal:  Br J Ophthalmol       Date:  2009-08-18       Impact factor: 4.638

5.  Inverse correlation between coronary and retinal blood flows in patients with normal coronary arteries and slow coronary blood flow.

Authors:  Yaron Arbel; Amir Sternfeld; Adiel Barak; Zvia Burgansky-Eliash; Amir Halkin; Shlomo Berliner; Itzhak Herz; Gad Keren; Ardon Rubinstein; Shmuel Banai; Ariel Finkelstein
Journal:  Atherosclerosis       Date:  2013-11-12       Impact factor: 5.162

6.  The correlation between retinal blood flow velocity measured by the retinal function imager and various physiological parameters.

Authors:  Zvia Burgansky-Eliash; Anat Lowenstein; Meira Neuderfer; Anat Kesler; Hila Barash; Darin A Nelson; Amiram Grinvald; Adiel Barak
Journal:  Ophthalmic Surg Lasers Imaging Retina       Date:  2013 Jan-Feb       Impact factor: 1.300

7.  Cilioretinal arteries in diabetic eyes are associated with increased retinal blood flow velocity and occurrence of diabetic macular edema.

Authors:  Gennady Landa; Wendewessen Amde; Yodit Haileselassie; Richard B Rosen
Journal:  Retina       Date:  2011-02       Impact factor: 4.256

Review 8.  Functional imaging using the retinal function imager: direct imaging of blood velocity, achieving fluorescein angiography-like images without any contrast agent, qualitative oximetry, and functional metabolic signals.

Authors:  David Izhaky; Darin A Nelson; Zvia Burgansky-Eliash; Amiram Grinvald
Journal:  Jpn J Ophthalmol       Date:  2009-09-08       Impact factor: 2.447

9.  Macular microcirculation in cystoid maculopathy of diabetic patients.

Authors:  O Arend; A Remky; A Harris; B Bertram; M Reim; S Wolf
Journal:  Br J Ophthalmol       Date:  1995-07       Impact factor: 4.638

10.  High-resolution wide-field imaging of perfused capillaries without the use of contrast agent.

Authors:  Darin A Nelson; Zvia Burgansky-Eliash; Hila Barash; Anat Loewenstein; Adiel Barak; Elisha Bartov; Tali Rock; Amiram Grinvald
Journal:  Clin Ophthalmol       Date:  2011-08-09
View more
  1 in total

1.  Comparison of Retinal Microvessel Blood Flow Velocities Acquired with Two Different Fields of View.

Authors:  Jin Zhou; Min Li; Wan Chen; Ye Yang; Liang Hu; Liang Wang; Hong Jiang; Jianhua Wang
Journal:  J Ophthalmol       Date:  2017-07-05       Impact factor: 1.909

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.