Literature DB >> 26372523

Influence of Different Tibial Fixation Techniques on Initial Stability in Single-Stage Anterior Cruciate Ligament Revision With Confluent Tibial Tunnels: A Biomechanical Laboratory Study.

Benedikt Schliemann1, Maximilian Treder1, Martin Schulze1, Viktoria Müller1, Sebastiano Vasta2, Biaggio Zampogna2, Mirco Herbort1, Clemens Kösters1, Michael J Raschke1, Simon Lenschow3.   

Abstract

PURPOSE: To kinematically and biomechanically compare 4 different types of tibial tunnel management in single-stage anterior cruciate ligament (ACL) revision reconstruction with the control: primary ACL reconstruction using a robotic-based knee testing setup.
METHODS: Porcine knees and flexor tendons were used. One hundred specimens were randomly assigned to 5 testing groups: (1) open tibial tunnel, (2) bone plug technique, (3) biodegradable interference screw, (4) dilatation technique, and (5) primary ACL reconstruction. A robotic/universal force-moment sensor testing system was used to simulate the KT-1000 (MEDmetric, San Diego, CA) and pivot-shift tests. Cyclic loading and load-to-failure testing were performed.
RESULTS: Anterior tibial translation increased significantly with all of the techniques compared with the intact ACL (P < .05). In the simulated KT-1000 test, groups 2 and 3 achieved results equal to those of primary ACL reconstruction (P > .05). The open tunnel and dilated tunnel techniques showed significantly greater anterior tibial translation (P < .05). The results of the simulated pivot-shift test were in accordance with those of the KT-1000 test. No significant differences could be observed regarding stiffness or maximum load to failure. However, elongation was significantly lower in the primary ACL reconstruction group compared with groups 1 and 3 (P = .02 and P = .03, respectively).
CONCLUSIONS: Filling an incomplete and incorrect tibial tunnel with a press-fit bone plug or a biodegradable interference screw in a standardized laboratory situation provided initial biomechanical properties and knee stability comparable with those of primary ACL reconstruction. In contrast, the dilatation technique or leaving the malplaced tunnel open did not restore knee kinematics adequately in this model. Backup extracortical fixation should be considered because the load to failure depends on the extracortical fixation when an undersized interference screw is used for aperture fixation. CLINICAL RELEVANCE: Our biomechanical results could help orthopaedic surgeons to optimize the results of primary ACL revision with incomplete, incorrect tunnel placement.
Copyright © 2016 Arthroscopy Association of North America. Published by Elsevier Inc. All rights reserved.

Mesh:

Year:  2015        PMID: 26372523     DOI: 10.1016/j.arthro.2015.06.044

Source DB:  PubMed          Journal:  Arthroscopy        ISSN: 0749-8063            Impact factor:   4.772


  3 in total

Review 1.  Revision ACL Reconstruction: Principles and Practice.

Authors:  Sachin Tapasvi; Anshu Shekhar
Journal:  Indian J Orthop       Date:  2021-01-19       Impact factor: 1.251

2.  Primary stability of single-stage revision reconstruction of the anterior cruciate ligament in case of failure of dynamic intraligamentary stabilization depends on implant position during ACL repair.

Authors:  B Schliemann; C Kösters; J Glasbrenner; M Fischer; M J Raschke; T Briese; M Müller; E Herbst; C Kittl
Journal:  Arch Orthop Trauma Surg       Date:  2021-07-31       Impact factor: 2.928

3.  Single-Stage Revision Anterior Cruciate Ligament Reconstruction Using the Stacked Screws Technique.

Authors:  Ian D Engler; Stephen M Sylvia; Matthew J Salzler; Julianne M Forlizzi; Thomas J Gill
Journal:  Arthrosc Tech       Date:  2022-06-21