| Literature DB >> 26368679 |
Nikolai N Klimov, Sunil Mittal, Michaela Berger, Zeeshan Ahmed.
Abstract
Resistance thermometry is a time-tested method for taking temperature measurements. In recent years, fundamental limits to resistance-based approaches have spurred considerable interest in developing photonic temperature sensors as a viable alternative. In this study, we demonstrate that our photonic thermometer, which consists of a silicon waveguide integrated with a Bragg grating, can be used to measure temperature changes over the range of 5°C-160°C, with a combined expanded uncertainty [k=2, 95% confidence level] of 1.25°C. Computational modeling of the sensor predicts the resonance wavelength and effective refractive index within 4% of the measured value.Year: 2015 PMID: 26368679 DOI: 10.1364/OL.40.003934
Source DB: PubMed Journal: Opt Lett ISSN: 0146-9592 Impact factor: 3.776