| Literature DB >> 26367964 |
Pedro Martín-Mateos, Borja Jerez, Pablo Acedo.
Abstract
In this paper, a multiheterodyne architecture for molecular dispersion spectroscopy based on a coherent dual-comb source generated using a single continuous wave laser and electro-optic modulators is presented and validated. The phase-sensitive scheme greatly simplifies previous dual-comb implementations by the use of an electro-optic dual comb and by phase-locking all the signal generators of the setup eliminating, in this way, the necessity of any reference optical path currently mandatory in absorption-based instruments. The architecture is immune to the classical baseline and normalization problems of absorption-based analyzers and provides an output linearly dependent on the gas concentration. In addition, the simultaneous parallel multi-wavelength measurement approach has the ability to deliver an improved output bandwidth (measurement speed) over gas analyzers based on tunable lasers.Year: 2015 PMID: 26367964 DOI: 10.1364/OE.23.021149
Source DB: PubMed Journal: Opt Express ISSN: 1094-4087 Impact factor: 3.894