Literature DB >> 26318084

Serum Metal Ion Levels Following Total Hip Arthroplasty With Modular Dual Mobility Components.

Laura J Matsen Ko1, Kimberley E Pollag1, Joanne Y Yoo1, Peter F Sharkey1.   

Abstract

Dual mobility acetabular components can reduce the incidence of total hip arthroplasty (THA) instability. Modular dual mobility (MDM) components facilitate acetabular component implantation. However, corrosion can occur at modular junctions. Serum cobalt and chromium levels and Oxford scores were obtained at minimum two year follow-up for 100 consecutive patients who had THA with MDM components. Average Oxford score was 43 (range 13-48). Average serum cobalt and chromium values were 0.7 mcg/L (range, 0.0 to 7.0) and 0.6 mcg/L (range, 0.1 to 2.7), respectively. MARS MRI was performed for four patients with pain and elevated serum cobalt levels. Two of these studies were consistent with adverse local tissue reaction. We recommend use of MDM implants in only patients at high risk for dislocation following THA.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  adverse local tissue reaction; chromium level; cobalt level; fretting corrosion; modular dual mobility components; total hip arthroplasty

Mesh:

Substances:

Year:  2015        PMID: 26318084     DOI: 10.1016/j.arth.2015.07.035

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  23 in total

1.  Revision for taper corrosion at the neck-body junction following total hip arthroplasty: pearls and pitfalls.

Authors:  Mitchell C Weiser; Darwin D Chen
Journal:  Curr Rev Musculoskelet Med       Date:  2016-03

2.  Understanding wear in dual mobility total hip replacement: first generation explant wear patterns.

Authors:  Bertrand Boyer; Thomas Neri; Jean Geringer; Alexandre Di Iorio; Remi Philippot; Frederic Farizon
Journal:  Int Orthop       Date:  2016-12-07       Impact factor: 3.075

Review 3.  The Role of One-Stage Exchange for Prosthetic Joint Infection.

Authors:  Fiachra E Rowan; Matthew J Donaldson; Jurek R Pietrzak; Fares S Haddad
Journal:  Curr Rev Musculoskelet Med       Date:  2018-09

4.  CORR Insights®: What Is the Survivorship of Revision Surgery Performed for the Chronically Dislocated THA?

Authors:  Denis Nam
Journal:  Clin Orthop Relat Res       Date:  2019-02       Impact factor: 4.176

5.  Bousquet dual mobility socket for patient under fifty years old. More than twenty year follow-up of one hundred and thirty one hips.

Authors:  Remi Philippot; Thomas Neri; Bertrand Boyer; Brice Viard; Frederic Farizon
Journal:  Int Orthop       Date:  2017-01-16       Impact factor: 3.075

6.  Lack of early dislocation for dual mobility vs. fixed bearing total hip arthroplasty: A multi-center analysis of comparable cohorts.

Authors:  J A Dubin; G H Westrich
Journal:  J Orthop       Date:  2020-02-04

Review 7.  Serum metal ion levels in modular dual mobility acetabular components: A systematic review.

Authors:  Ioannis Gkiatas; Abhinav K Sharma; Alexander Greenberg; Stephen T Duncan; Brian P Chalmers; Peter K Sculco
Journal:  J Orthop       Date:  2020-08-25

Review 8.  Metal ion levels with use of modular dual mobility constructs: Can the evidence guide us on clinical use?

Authors:  Atul F Kamath; P Maxwell Courtney; Gwo-Chin Lee
Journal:  J Orthop       Date:  2021-02-20

Review 9.  Adverse reaction to metal debris due to fretting corrosion between the acetabular components of modular dual-mobility constructs in total hip replacement: a systematic review and meta-analysis.

Authors:  Jonathan M R French; Paul Bramley; Sean Scattergood; Nemandra A Sandiford
Journal:  EFORT Open Rev       Date:  2021-05-04

10.  Severe Corrosion of Modular Dual Mobility Acetabular Components Identified During Revision Total Hip Arthroplasty.

Authors:  Mohammad S Abdelaal; Eric Zachwieja; Peter F Sharkey
Journal:  Arthroplast Today       Date:  2021-03-03
View more