Literature DB >> 26316899

Development of A Physical Windkessel Module to Re-Create In-Vivo Vascular Flow Impedance for In-Vitro Experiments.

Ethan O Kung1, Charles A Taylor2.   

Abstract

PURPOSE: To create and characterize a physical Windkessel module that can provide realistic and predictable vascular impedances for in-vitro flow experiments used for computational fluid dynamics validation, and other investigations of the cardiovascular system and medical devices.
METHODS: We developed practical design and manufacturing methods for constructing flow resistance and capacitance units. Using these units we assembled a Windkessel impedance module and defined its corresponding analytical model incorporating an inductance to account for fluid momentum. We tested various resistance units and Windkessel modules using a flow system, and compared experimental measurements to analytical predictions of pressure, flow, and impedance.
RESULTS: The resistance modules exhibited stable resistance values over wide ranges of flow rates. The resistance value variations of any particular resistor are typically within 5% across the range of flow that it is expected to accommodate under physiologic flow conditions. In the Windkessel impedance modules, the measured flow and pressure waveforms agreed very favorably with the analytical calculations for four different flow conditions used to test each module. The shapes and magnitudes of the impedance modulus and phase agree well between experiment and theoretical values, and also with those measured in-vivo in previous studies.
CONCLUSIONS: The Windkessel impedance module we developed can be used as a practical tool to provide realistic vascular impedance for in-vitro cardiovascular studies. Upon proper characterization of the impedance module, its analytical model can accurately predict its measured behavior under different flow conditions.

Entities:  

Keywords:  Boundary condition; Flow capacitance; Flow impedance; Flow inductance; Flow resistance; Windkessel; blood flow; in-vitro validation; vascular impedance

Year:  2011        PMID: 26316899      PMCID: PMC4548962          DOI: 10.1007/s13239-010-0030-6

Source DB:  PubMed          Journal:  Cardiovasc Eng Technol        ISSN: 1869-408X            Impact factor:   2.495


  16 in total

1.  Pulmonary arterial compliance in dogs and pigs: the three-element windkessel model revisited.

Authors:  P Segers; S Brimioulle; N Stergiopulos; N Westerhof; R Naeije; M Maggiorini; P Verdonck
Journal:  Am J Physiol       Date:  1999-08

2.  A fractal continuum model of the pulmonary arterial tree.

Authors:  G S Krenz; J H Linehan; C A Dawson
Journal:  J Appl Physiol (1985)       Date:  1992-06

3.  Comparison of CFD and MRI flow and velocities in an in vitro large artery bypass graft model.

Authors:  Joy P Ku; Christopher J Elkins; Charles A Taylor
Journal:  Ann Biomed Eng       Date:  2005-03       Impact factor: 3.934

4.  Renal blood flow measurements with use of phase-contrast magnetic resonance imaging: normal values and reproducibility.

Authors:  Liesbeth Bax; Chris J G Bakker; Willemijn M Klein; Niels Blanken; Jaap J Beutler; Willem P T R M Mali
Journal:  J Vasc Interv Radiol       Date:  2005-06       Impact factor: 3.464

5.  Systemic venous circulation. Waves propagating on a windkessel: relation of arterial and venous windkessels to systemic vascular resistance.

Authors:  Jiun-Jr Wang; Jacqueline A Flewitt; Nigel G Shrive; Kim H Parker; John V Tyberg
Journal:  Am J Physiol Heart Circ Physiol       Date:  2005-08-19       Impact factor: 4.733

6.  Fractal network model for simulating abdominal and lower extremity blood flow during resting and exercise conditions.

Authors:  Brooke N Steele; Mette S Olufsen; Charles A Taylor
Journal:  Comput Methods Biomech Biomed Engin       Date:  2007-02       Impact factor: 1.763

7.  An artificial arterial system for pumping hearts.

Authors:  N Westerhof; G Elzinga; P Sipkema
Journal:  J Appl Physiol       Date:  1971-11       Impact factor: 3.531

8.  Characterization of pulmonary arterial input impedance with lumped parameter models.

Authors:  B J Grant; L J Paradowski
Journal:  Am J Physiol       Date:  1987-03

9.  Quantification of hemodynamics in abdominal aortic aneurysms during rest and exercise using magnetic resonance imaging and computational fluid dynamics.

Authors:  Andrea S Les; Shawn C Shadden; C Alberto Figueroa; Jinha M Park; Maureen M Tedesco; Robert J Herfkens; Ronald L Dalman; Charles A Taylor
Journal:  Ann Biomed Eng       Date:  2010-02-09       Impact factor: 3.934

10.  Noninvasive characterization of renal artery blood flow.

Authors:  E R Greene; M D Venters; P S Avasthi; R L Conn; R W Jahnke
Journal:  Kidney Int       Date:  1981-10       Impact factor: 10.612

View more
  10 in total

1.  A New Multi-Mode Perfusion System for Ex Vivo Heart Perfusion Study.

Authors:  Liming Xin; Bryan Gellner; Roberto Vanin Pinto Ribeiro; Giulia Maria Ruggeri; David Banner; Massimiliano Meineri; Vivek Rao; Jean Zu; Mitesh V Badiwala
Journal:  J Med Syst       Date:  2017-12-23       Impact factor: 4.460

2.  In vitro validation of finite-element model of AAA hemodynamics incorporating realistic outlet boundary conditions.

Authors:  Ethan O Kung; Andrea S Les; Francisco Medina; Ryan B Wicker; Michael V McConnell; Charles A Taylor
Journal:  J Biomech Eng       Date:  2011-04       Impact factor: 2.097

3.  In vitro validation of finite element analysis of blood flow in deformable models.

Authors:  Ethan O Kung; Andrea S Les; C Alberto Figueroa; Francisco Medina; Karina Arcaute; Ryan B Wicker; Michael V McConnell; Charles A Taylor
Journal:  Ann Biomed Eng       Date:  2011-03-15       Impact factor: 3.934

4.  A Zero-Dimensional Model and Protocol for Simulating Patient-Specific Pulmonary Hemodynamics From Limited Clinical Data.

Authors:  Vitaly O Kheyfets; Jamie Dunning; Uyen Truong; Dunbar Ivy; Kendall Hunter; Robin Shandas
Journal:  J Biomech Eng       Date:  2016-12-01       Impact factor: 2.097

Review 5.  Considerations for numerical modeling of the pulmonary circulation--a review with a focus on pulmonary hypertension.

Authors:  V O Kheyfets; W O'Dell; T Smith; J J Reilly; E A Finol
Journal:  J Biomech Eng       Date:  2013-06       Impact factor: 2.097

6.  Mock circulatory system of the Fontan circulation to study respiration effects on venous flow behavior.

Authors:  Marija Vukicevic; John A Chiulli; Timothy Conover; Giancarlo Pennati; Tain Yen Hsia; Richard S Figliola
Journal:  ASAIO J       Date:  2013 May-Jun       Impact factor: 2.872

7.  Pulse Wave Imaging Coupled With Vector Flow Mapping: A Phantom, Simulation, and In Vivo Study.

Authors:  Grigorios Marios Karageorgos; Iason-Zacharias Apostolakis; Pierre Nauleau; Vittorio Gatti; Rachel Weber; Paul Kemper; Elisa E Konofagou
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2021-06-29       Impact factor: 3.267

8.  A structural approach to 3D-printing arterial phantoms with physiologically comparable mechanical characteristics: Preliminary observations.

Authors:  Bruce Guest; Luis Arroyo; John Runciman
Journal:  Proc Inst Mech Eng H       Date:  2022-08-01       Impact factor: 1.763

9.  A novel nonlinear afterload for ex vivo heart evaluation: Porcine experimental results.

Authors:  Henry Pigot; Kristian Soltesz; Audrius Paskevicius; Qiuming Liao; Trygve Sjöberg; Stig Steen
Journal:  Artif Organs       Date:  2022-05-20       Impact factor: 2.663

10.  A new noninvasive and patient-specific hemodynamic index for the severity of renal stenosis and outcome of interventional treatment.

Authors:  Huidan Yu; Monsurul Khan; Hao Wu; Xiaoping Du; Rou Chen; Dave M Rollins; Xin Fang; Jianyun Long; Chenke Xu; Alan P Sawchuk
Journal:  Int J Numer Method Biomed Eng       Date:  2022-05-18       Impact factor: 2.648

  10 in total