Allergic Reaction following Implantation of a Nitinol Alloy Inferior Vena Cava Filter.

Zhongzhi Jia1, JianFei Tu2, Kai Wang1, Guomin Jiang1, Weiping Wang3.   

Abstract

A 67-year-old woman developed severe pruritus after implantation of a retrievable nitinol inferior vena cava (IVC) filter (55.4% nickel and 44.6% titanium). The pruritus resolved only after filter retrieval. The patient's hypersensitivity to nickel was confirmed by a positive skin patch test substantiating a systemic allergic reaction to the implanted nitinol IVC filter.
Copyright © 2015 SIR. Published by Elsevier Inc. All rights reserved.