Limitation of the activated partial thromboplastin time as a monitoring method of the direct thrombin inhibitor argatroban.

S Guy1, S Kitchen1, R Maclean1, J J Van Veen1.   

Abstract

BACKGROUND: Argatroban is licensed for patients with heparin-induced thrombocytopenia and monitoring is conventionally by activated partial thromboplastin time (APTT) ratio with a target of 1.5-3.0 and not exceeding 100 s. The APTT may be influenced by coagulopathies, lupus anticoagulant and raised FVIII levels. Variable but not clinically significant sensitivity of APTT reagents to argatroban has been highlighted in other studies.
METHODS: Residual plasma of 15 patients (n = 124 samples) was tested on Sysmex(™) CS series using the Hemoclot(®) thrombin inhibitor assay (HTI) and APTT ratio (Actin FS and SynthASil). A subgroup from four patients (n = 31) were tested on ACL TOP(™) to compare the different platforms with the HTI. Spiked normal pooled plasma was tested with Actin FS, Actin FSL, SynthASil and APTT-SP on their respective platforms (CS5100(™) and ACL TOP(™)) to assess reagent sensitivity.
RESULTS: Mean concentration of argatroban by HTI assay for patient plasma was 0.47 μg/mL; the mean APTT ratio using Actin FS was 1.89 and for SynthASil 1.56. There was a poor correlation between APTT and the HTI. In the spiked normal pooled plasma, Actin FS gave a significantly higher APTT ratio than the other three reagents for the various argatroban concentrations.
CONCLUSIONS: Hemoclot(®) thrombin inhibitor assay should be considered in patients on argatroban, particularly if there is concern the APTT may not be reflective of the degree of anticoagulation with argatroban due to other factors including coagulopathies in critically ill patients, the presence of a lupus anticoagulant or very high FVIII levels.
© 2015 John Wiley & Sons Ltd.

Entities:  

Keywords:  Argatroban; Hemoclot thrombin inhibitor assay; activated partial thromboplastin time; heparin-induced thrombocytopenia; monitoring

Mesh:

Substances:

Year:  2015        PMID: 26305560     DOI: 10.1111/ijlh.12414

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  4 in total

1.  Evaluation of intravenous direct thrombin inhibitor monitoring tests: Correlation with plasma concentrations and clinical outcomes in hospitalized patients.

Authors:  Jacob T Beyer; Stuart E Lind; Sheila Fisher; Toby C Trujillo; Michael F Wempe; Tyree H Kiser
Journal:  J Thromb Thrombolysis       Date:  2020-02       Impact factor: 2.300

2.  Monitoring of argatroban and lepirudin anticoagulation in critically ill patients by conventional laboratory parameters and rotational thromboelastometry - a prospectively controlled randomized double-blind clinical trial.

Authors:  Martin Beiderlinden; Patrick Werner; Astrid Bahlmann; Johann Kemper; Tobias Brezina; Maximilian Schäfer; Klaus Görlinger; Holger Seidel; Peter Kienbaum; Tanja A Treschan
Journal:  BMC Anesthesiol       Date:  2018-02-09       Impact factor: 2.217

3.  Caution in Using the Activated Partial Thromboplastin Time to Monitor Argatroban in COVID-19 and Vaccine-Induced Immune Thrombocytopenia and Thrombosis (VITT).

Authors:  Susan Guy; Steve Kitchen; Michael Makris; Rhona M Maclean; Giorgia Saccullo; Joost J Vanveen
Journal:  Clin Appl Thromb Hemost       Date:  2021 Jan-Dec       Impact factor: 2.389

Review 4.  "In Less than No Time": Feasibility of Rotational Thromboelastometry to Detect Anticoagulant Drugs Activity and to Guide Reversal Therapy.

Authors:  Vittorio Pavoni; Lara Gianesello; Duccio Conti; Piercarlo Ballo; Pietro Dattolo; Domenico Prisco; Klaus Görlinger
Journal:  J Clin Med       Date:  2022-03-04       Impact factor: 4.241

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.