| Literature DB >> 26300239 |
S Alibi1, A Ferjani2, O Gaillot2, M Marzouk2, R Courcol3, J Boukadida2.
Abstract
We evaluated the Bruker Biotyper matrix-assisted laser desorption ionization-time of flight (MALDI-TOF) mass spectrometry (MS) for the identification of 97 Corynebacterium clinical in comparison to identification strains by Api Coryne and MALDI-TOF-MS using 16S rRNA gene and hypervariable region of rpoB genes sequencing as a reference method. C. striatum was the predominant species isolated followed by C. amycolatum. There was an agreement between Api Coryne strips and MALDI-TOF-MS identification in 88.65% of cases. MALDI-TOF-MS was unable to differentiate C. aurimucosum from C. minutissimum and C. minutissimum from C. singulare but reliably identify 92 of 97 (94.84%) strains. Two strains remained incompletely identified to the species level by MALDI-TOF-MS and molecular approaches. They belonged to Cellulomonas and Pseudoclavibacter genus. In conclusion, MALDI-TOF-MS is a rapid and reliable method for the identification of Corynebacterium species. However, some limits have been noted and have to be resolved by the application of molecular methods.Entities:
Keywords: 16S rRNA gene sequencing; 16S rRNA; Api Coryne; Corynebacterium; Corynebacterium spp.; Identification; MALDI-TOF-MS; SM-MALDI-TOF; Séquençage; rpoB
Mesh:
Substances:
Year: 2015 PMID: 26300239 DOI: 10.1016/j.patbio.2015.07.007
Source DB: PubMed Journal: Pathol Biol (Paris) ISSN: 0369-8114