Literature DB >> 26294009

Fluid-Structure Interaction Model of a Percutaneous Aortic Valve: Comparison with an In Vitro Test and Feasibility Study in a Patient-Specific Case.

Wei Wu1, Desiree Pott2, Beniamino Mazza1, Tommaso Sironi1, Elena Dordoni1, Claudio Chiastra1,3, Lorenza Petrini4, Giancarlo Pennati1, Gabriele Dubini1, Ulrich Steinseifer2, Simon Sonntag2, Maximilian Kuetting2, Francesco Migliavacca5.   

Abstract

Transcatheter aortic valve replacement (TAVR) represents an established recent technology in a high risk patient base. To better understand TAVR performance, a fluid-structure interaction (FSI) model of a self-expandable transcatheter aortic valve was proposed. After an in vitro durability experiment was done to test the valve, the FSI model was built to reproduce the experimental test. Lastly, the FSI model was used to simulate the virtual implant and performance in a patient-specific case. Results showed that the leaflet opening area during the cycle was similar to that of the in vitro test and the difference of the maximum leaflet opening between the two methodologies was of 0.42%. Furthermore, the FSI simulation quantified the pressure and velocity fields. The computed strain amplitudes in the stent frame showed that this distribution in the patient-specific case is highly affected by the aortic root anatomy, suggesting that the in vitro tests that follow standards might not be representative of the real behavior of the percutaneous valve. The patient-specific case also compared in vivo literature data on fast opening and closing characteristics of the aortic valve during systolic ejection. FSI simulations represent useful tools in determining design errors or optimization potentials before the fabrication of aortic valve prototypes and the performance of tests.

Entities:  

Keywords:  Fluid–structure interaction; Mathematical models; Stent; Transcatheter aortic valve; Valve mechanics

Mesh:

Year:  2015        PMID: 26294009     DOI: 10.1007/s10439-015-1429-x

Source DB:  PubMed          Journal:  Ann Biomed Eng        ISSN: 0090-6964            Impact factor:   3.934


  22 in total

1.  Realistic Vascular Replicator for TAVR Procedures.

Authors:  Oren M Rotman; Brandon Kovarovic; Chander Sadasivan; Luis Gruberg; Baruch B Lieber; Danny Bluestein
Journal:  Cardiovasc Eng Technol       Date:  2018-04-13       Impact factor: 2.495

2.  Immersogeometric cardiovascular fluid-structure interaction analysis with divergence-conforming B-splines.

Authors:  David Kamensky; Ming-Chen Hsu; Yue Yu; John A Evans; Michael S Sacks; Thomas J R Hughes
Journal:  Comput Methods Appl Mech Eng       Date:  2016-08-04       Impact factor: 6.756

3.  Toward the Virtual Benchmarking of Pneumatic Ventricular Assist Devices: Application of a Novel Fluid-Structure Interaction-Based Strategy to the Penn State 12 cc Device.

Authors:  Alessandro Caimi; Francesco Sturla; Bryan Good; Marco Vidotto; Rachele De Ponti; Filippo Piatti; Keefe B Manning; Alberto Redaelli
Journal:  J Biomech Eng       Date:  2017-08-01       Impact factor: 2.097

Review 4.  Principles of TAVR valve design, modelling, and testing.

Authors:  Oren M Rotman; Matteo Bianchi; Ram P Ghosh; Brandon Kovarovic; Danny Bluestein
Journal:  Expert Rev Med Devices       Date:  2018-10-29       Impact factor: 3.166

5.  Fluid-Structure Interaction Study of Transcatheter Aortic Valve Dynamics Using Smoothed Particle Hydrodynamics.

Authors:  Wenbin Mao; Kewei Li; Wei Sun
Journal:  Cardiovasc Eng Technol       Date:  2016-11-14       Impact factor: 2.495

Review 6.  Computational Fluid Dynamics Assessment Associated with Transcatheter Heart Valve Prostheses: A Position Paper of the ISO Working Group.

Authors:  Zhenglun Alan Wei; Simon Johannes Sonntag; Milan Toma; Shelly Singh-Gryzbon; Wei Sun
Journal:  Cardiovasc Eng Technol       Date:  2018-04-19       Impact factor: 2.495

7.  Numerical evaluation of transcatheter aortic valve performance during heart beating and its post-deployment fluid-structure interaction analysis.

Authors:  Ram P Ghosh; Gil Marom; Matteo Bianchi; Karl D'souza; Wojtek Zietak; Danny Bluestein
Journal:  Biomech Model Mechanobiol       Date:  2020-02-24

Review 8.  Biomechanical Behavior of Bioprosthetic Heart Valve Heterograft Tissues: Characterization, Simulation, and Performance.

Authors:  Joao S Soares; Kristen R Feaver; Will Zhang; David Kamensky; Ankush Aggarwal; Michael S Sacks
Journal:  Cardiovasc Eng Technol       Date:  2016-08-09       Impact factor: 2.495

9.  Immersed Methods for Fluid-Structure Interaction.

Authors:  Boyce E Griffith; Neelesh A Patankar
Journal:  Annu Rev Fluid Mech       Date:  2019-09-05       Impact factor: 18.511

10.  The Comparison of Different Constitutive Laws and Fiber Architectures for the Aortic Valve on Fluid-Structure Interaction Simulation.

Authors:  Li Cai; Ruihang Zhang; Yiqiang Li; Guangyu Zhu; Xingshuang Ma; Yongheng Wang; Xiaoyu Luo; Hao Gao
Journal:  Front Physiol       Date:  2021-06-24       Impact factor: 4.566

View more