Sudden and Fatal Malfunction of a Durata Defibrillator Lead due to External Insulation Failure.

Anand D Shah1, David S Hirsh1, Jonathan J Langberg1.   

Abstract

Defibrillator lead malfunction can be a disastrous complication, leading to loss of protection from sudden cardiac death in a high-risk patient population. Recognition of lead-specific risk for failure can assist in development of focused screening or surveillance, as in the case of the Riata lead (St. Jude Medical, St. Paul, MN, USA) or the Sprint Fidelis lead (Medtronic Inc., Minneapolis, MN, USA). A case of defibrillation failure secondary to a Durata lead insulation failure is presented. A brief review of the literature and current St. Jude Medical implantable cardiac defibrillator lead design is presented. Identification of arcing is identified as a potential sign of catastrophic insulation failure. ©2015 Wiley Periodicals, Inc.

Entities:  

Keywords:  Durata; ICD failure; abrasion; insulation; optim

Mesh:

Year:  2015        PMID: 26293817     DOI: 10.1111/pace.12738

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  3 in total

1.  Defibrillation failure with short ICD charge time: Internal short circuit of a Durata lead.

Authors:  Ayako Okada; Wataru Shoin; Morio Shoda; Hiroaki Tabata; Hideki Kobayashi; Takahiro Okano; Koji Yoshie; Ken Kato; Koichiro Kuwahara
Journal:  J Cardiol Cases       Date:  2019-10-31

2.  Unusual fracture in a Durata lead with shock coil fragmentation and cable externalization.

Authors:  Toshiaki Sato; Kyoko Soejima; Hideaki Yoshino; Morio Shoda
Journal:  HeartRhythm Case Rep       Date:  2017-05-26

3.  A Primer to the Structure, Content and Linkage of the FDA's Manufacturer and User Facility Device Experience (MAUDE) Files.

Authors:  Lisa Garnsey Ensign; K Bretonnel Cohen
Journal:  EGEMS (Wash DC)       Date:  2017-06-14
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.