Literature DB >> 26291952

Separating the Laparoscopic Camera Cord From the Monopolar "Bovie" Cord Reduces Unintended Thermal Injury From Antenna Coupling: A Randomized Controlled Trial.

Thomas N Robinson1, Edward L Jones, Christina L Dunn, Bruce Dunne, Elizabeth Johnson, Nicole T Townsend, Alessandro Paniccia, Greg V Stiegmann.   

Abstract

OBJECTIVE(S): The monopolar "Bovie" is used in virtually every laparoscopic operation. The active electrode and its cord emit radiofrequency energy that couples (or transfers) to nearby conductive material without direct contact. This phenomenon is increased when the active electrode cord is oriented parallel to another wire/cord. The parallel orientation of the "Bovie" and laparoscopic camera cords cause transfer of energy to the camera cord resulting in cutaneous burns at the camera trocar incision. We hypothesized that separating the active electrode/camera cords would reduce thermal injury occurring at the camera trocar incision in comparison to parallel oriented active electrode/camera cords.
METHODS: In this prospective, blinded, randomized controlled trial, patients undergoing standardized laparoscopic cholecystectomy were randomized to separated active electrode/camera cords or parallel oriented active electrode/camera cords. The primary outcome variable was thermal injury determined by histology from skin biopsied at the camera trocar incision.
RESULTS: Eighty-four patients participated. Baseline demographics were similar in the groups for age, sex, preoperative diagnosis, operative time, and blood loss. Thermal injury at the camera trocar incision was lower in the separated versus parallel group (31% vs 57%; P = 0.027).
CONCLUSIONS: Separation of the laparoscopic camera cord from the active electrode cord decreases thermal injury from antenna coupling at the camera trocar incision in comparison to the parallel orientation of these cords. Therefore, parallel orientation of these cords (an arrangement promoted by integrated operating rooms) should be abandoned. The findings of this study should influence the operating room setup for all laparoscopic cases.

Entities:  

Mesh:

Year:  2015        PMID: 26291952     DOI: 10.1097/SLA.0000000000000841

Source DB:  PubMed          Journal:  Ann Surg        ISSN: 0003-4932            Impact factor:   12.969


  6 in total

1.  Stray energy transfer in single-incision robotic surgery.

Authors:  Krzysztof J Wikiel; Douglas M Overbey; Heather Carmichael; Brandon C Chapman; John T Moore; Carlton C Barnett; Teresa S Jones; Thomas N Robinson; Edward L Jones
Journal:  Surg Endosc       Date:  2020-06-26       Impact factor: 4.584

2.  Single-incision laparoscopic surgery increases the risk of unintentional thermal injury from the monopolar "Bovie" instrument in comparison with traditional laparoscopy.

Authors:  Nicole T Townsend; Edward L Jones; Doug Overbey; Bruce Dunne; Jennifer McHenry; Thomas N Robinson
Journal:  Surg Endosc       Date:  2016-11-18       Impact factor: 4.584

3.  The SAGES Fundamental Use of Surgical Energy program (FUSE): history, development, and purpose.

Authors:  P Fuchshuber; S Schwaitzberg; D Jones; S B Jones; L Feldman; M Munro; T Robinson; G Purcell-Jackson; D Mikami; A Madani; M Brunt; B Dunkin; C Gugliemi; L Groah; R Lim; J Mischna; C R Voyles
Journal:  Surg Endosc       Date:  2017-12-07       Impact factor: 4.584

4.  Monopolar stray energy in robotic surgery.

Authors:  Douglas M Overbey; Heather Carmichael; Krzysztof J Wikiel; Douglas A Hirth; Brandon C Chapman; John T Moore; Carlton C Barnett; Teresa S Jones; Thomas N Robinson; Edward L Jones
Journal:  Surg Endosc       Date:  2020-05-08       Impact factor: 4.584

5.  Robotic stray energy with constant-voltage versus constant-power regulating electrosurgical generators.

Authors:  Krzysztof J Wikiel; Franklin J Powlan; Teresa S Jones; Thomas N Robinson; Edward L Jones
Journal:  Surg Endosc       Date:  2022-05-25       Impact factor: 4.584

6.  Comparing the safety and effectiveness of dedicated radiofrequency transseptal wires to electrified metal guidewires.

Authors:  Jeremiah Wasserlauf; Bradley P Knight
Journal:  J Cardiovasc Electrophysiol       Date:  2022-01-11       Impact factor: 2.942

  6 in total